87
Status Commands
∗
CLS
Clear status. Clears the Standard Event Status Register, Query
Error Register and Execution Error Register. This indirectly clears
the Status Byte Register.
∗
ESE <nrf>
Set the Standard Event Status Enable Register to the value of
<nrf>.
∗
ESE?
Returns the value in the Standard Event Status Enable Register
in <nr1> numeric format. The syntax of the response is
<nr1><rmt>.
∗
ESR?
Returns the value in the Standard Event Status Register in <nr1>
numeric format. The register is then cleared. The syntax of the
response is <nr1><rmt>.
∗
IDN?
Returns the instrument identification. The exact response is
determined by the instrument configuration and is of the form
<NAME>, <model>, 0, <version><rmt>where <NAME> is the
manufacturer’s name, <MODEL> defines the type of instrument
and <VERSION> is the revision level of the software installed.
∗
IST?
Returns
ist
local message as defined by IEEE Std. 488.2. The
syntax of the response if 0<rmt>, if the local message false or
1<rmt>, if the local message is true.
∗
OPC
Sets the Operation Complete bit (bit 0) in the Standard Event
Status Register. This will happen immediately the command is
executed because of the sequential nature of all operations.
∗
OPC?
Query operation complete status. The syntax of the response is
1<rmt>. The response will be available immediately the
command is executed because of the sequential nature of all
operations.
∗
PRE <nrf>
Set the Parallel Poll Enable Register to the value <nrf>.
∗
PRE?
Returns the value in the Parallel Poll Enable Register in <nr1>
numeric format. The syntax of the response is <nr1><rmt>.
∗
SRE <nrf>
Set the Service Request Enable Register to <nrf>.
∗
SRE?
Returns the value of the Service Request Enable Register in
<nr1> numeric format. The Syntax of the response is
<nr1><rmt>.
∗
STB?
Returns the value of the Status Byte Register in <nr1> numeric
format. The syntax of the response is <nr1><rmt>.
∗
WAI
Wait for operation complete true. As all commands are
completely executed before the next is started this command
takes no additional action.
∗
TST?
The generator has no self
−
test capability and the response is
always 0<rmt>.
EER?
Query and clear execution error number register. The response
format is nr1<rmt>.
QER?
Query and clear query error number register. The response
format is nr1<rmt>.