14
Fuse
Ensure that the correct mains fuse is fitted for the set operating voltage. The correct mains fuse
types are:
Single channel
for 230V operation:
250 mA (T) 250V HRC
for 100V or 115V operation:
500 mA (T) 250V HRC
2 & 4 channel
for 230V operation:
1A(T) 250V HRC
for 100V or 115V operation:
2A(T) 250V HRC
To replace the fuse, disconnect the mains lead from the inlet socket and withdraw the fuse
drawer below the socket pins. Change the fuse and replace the drawer.
The use of makeshift fuses or the short
−
circuiting of the fuse holder is prohibited.
Mains Lead
Connect the instrument to the AC supply using the mains lead provided. Should a mains plug be
required for a different mains outlet socket, a suitably rated and approved mains lead set should
be used which is fitted with the required wall plug and an IEC60320 C13 connector for the
instrument end. To determine the minimum current rating of the lead-set for the intended AC
supply, refer to the power rating information on the equipment or in the Specification.
WARNING! THIS INSTRUMENT MUST BE EARTHED
Any interruption of the mains earth conductor inside or outside the instrument will make the
instrument dangerous. Intentional interruption is prohibited. The protective action must not be
negated by the use of an extension cord without a protective conductor.
Mounting
This instrument is suitable both for bench use and rack mounting. It is delivered with feet for
bench mounting. The front feet include a tilt mechanism for optimal panel angle.
A rack kit for mounting in a 19” rack is available from the Manufacturers or their overseas agents.