4
DISPLAY FEATURES
The display is a 256 x 112 pixel transflective LCD, backlit by white LEDs; contrast is software-
controlled and can be adjusted from the keyboard.
The default status display shows the VOLTS (5 digits) and AMPS (4 digits) in 12mm high
characters, plus the present output mode. At other times, for example during store or recall of
instrument set-ups, the display shows up to 7 lines of information, instructions, or prompts.
INTERFACES (QPX1200SP only)
Full digital remote control facilities are available through the RS232, LAN, USB and GPIB
interfaces. Setting and readback resolutions are the same as the Output and Meter specifications
respectively.
RS232:
Standard 9-pin D-connector. Baud rate 9600.
USB:
Standard USB 2.0 hardware connection.
LAN:
Ethernet 100/10base-T hardware connection.
1.4 LXI Core 2011 compliant.
GPIB:
Conforming with IEEE488.1 and IEEE488.2.
Remote Command
Processing Time:
Typically <100ms between receiving the command terminator for a
step voltage change at the instrument and the output voltage
beginning to change.
GENERAL
AC Input:
110V – 240V AC ± 10%, 50/60Hz.
Installation Category II.
Power Consumption:
1600VA max.
Operating Range:
+5ºC to +40ºC, 20% to 80% RH.
Storage Range:
−
40ºC to + 70ºC.
Environmental:
Indoor use at altitudes up to 2000m, Pollution Degree 2.
Store/Recall:
Up to 10 set-ups can be saved and recalled via the keyboard or
remote interfaces.
Safety:
Complies with EN61010-1 & EN61326-1.
For details, request the EU Declaration of Conformity for this
instrument via http://www.aimtti.com/support (serial no. needed).
Size:
130mm H (3U) x 356mm W x 413mm D
Weight:
9.2kg
Options:
19-inch rack kit.