TROUBLESHOOTING
Error message / Troubleshooting
Possible cause
Corrective action
“Cuvette position not correct.
Remove cuvette.” error message is displayed.
Remnants of cuvette sealing foil is left on
the cuvette collar.
Remove cuvette when the instrument has lifted
it up. Make sure that all remnants are removed
when doing the next measurement.
The instrument has a mechanical malfunction.
Check the item above. If it does not apply,
restart the instrument. If the problem persists,
contact customer service.
Measurement prohibited.
Reagent cap is missing or cuvette is used.
Check that the cuvette has a reagent cap and
the inner coloured part of the cap is not pressed
down.
Reading lot data from the barcode failed.
Try again. If the problem persists, cancel the
test.
Kit lot expired.
Discard expired kit lot. Use a new one.
Cuvette temperature too low.
Let the cuvette warm up to room temperature.
Test the same cuvette again.
Cuvette temperature too high.
Let the cuvette cool down to room temperature.
Test the same cuvette again.
Test cancelled.
Blank too high.
Test the same cuvette again. The blanking
process has not been completed or the
sample may contain interfering substances.
In the latter case, the test cannot be completed.
Unstable blank.
Error in reagent addition.
Perform a new test. There has been some
problem during reagent addition. Make sure that
the cap is properly closed.
Instrument failure.
Perform a new test. If this message appears
often, contact customer service.