Remove Cap, fill Bulb full of warm soapy water, replace Cap
and force water through Tubing by gently squeezing the Bulb.
Remove Cap, refill with clean water, replace Cap and again,
gently squeeze the Bulb to force water through the Tubing.
Completely disassemble the FingerFeeder and clean all parts
(except Tape) with warm soapy water.
Rinse parts well and allow to air dry on a clean towel. (It may
help to reassemble the device and gently squeeze the empty
Bulb a few times to remove water droplets from the Tubing).
1.
2.
3.
4.
Cleaning Instructions
After each use:
Once per day:
This device may be sterilized or sanitized by using the following
instructions.
Sterilization:
All disassembled parts can be autoclaved to a maximum of 272°
for 3 minutes at 29 PSIG or 250° F for 15 minutes at 15 PSIG.
To avoid damage during and immediately after sterilization:
• Sterilize without applying external pressure on parts.
• Allow parts to cool before applying external pressure.
If gas sterilization is used, residual testing is recommended to
determine the proper length of aeration.
Sanitization:
• Place parts in a 4-5 quart pan.
• Fill to within 1inch of rim with cold water. Distilled water is preferred, to reduce hard water deposit
formation.
• Cover pan and bring water to a boil.
• Boil parts for 20 minutes.
• Remove pan from heat.
• Drain water and allow parts to cool in open pan.
• Remove all parts and allow to air dry on a clean towel.
• Avoid touching the inside of the Bulb.