1) Constant display mode
During normal operation (while a flow rate value is displayed), pressing and holding the
“FLOW RATE” switch and the “START” switch together for four seconds or longer will
change the display to this mode.
You can select various setting items in sequence by pressing the “FLOW RATE” switch.
Pressing the “FLOW RATE” switch for two seconds or longer at the item you want to set
up (configurable constant only) allows you to set up the value.
Press the “FLOW RATE” switch to select the setting, and press and hold the “FLOW
RATE” switch again for two seconds or longer to confirm the setting.
When the “START” switch is pressed or no switch operations have been done for ten
seconds or longer during the selection of setting items, the display will return to an
integrated flow rate value.
When no input operations have been done for ten seconds or longer during setting, the
display will return to an integrated flow rate value even if the setting has not been
completed.
When the setting has not been completed, the constant data to be used is the previous
setting value.
The following figure shows the flow of constant setting and display.
Constant display mode: Press and hold the “FLOW RATE” the “START” switch for four seconds
1. Serial number display
Press the “FLOW RATE” switch.
2. Integrated pulse constant display
Press the “FLOW RATE” switch.
3. High-density pulse output constant display
Press the “FLOW RATE” switch.
4. Pulse output unit setting
5. Pulse output width setting
Press the “FLOW RATE” switch.
Press and hold the “FLOW RATE”
switch for two seconds
(Start of setting)
Pulse output unit is being set
Press and hold the “FLOW
RATE” switch for two seconds
(Confirmation of setting)
Press the “FLOW RATE” switch.
Press and hold the “FLOW RATE”
switch for two seconds
(Start of setting)
Press and hold the “FLOW
RATE” switch for two seconds
(Confirmation of setting)
Pulse output width is being set
When the
“START” switch is
pressed or no switch
operations have been
done for ten seconds or
longer during the
selection of setting
items, or when no
switch operations have
not been done for ten
seconds or longer
during the setting of an
item, the display will
switch to an integrated
flow rate value.