Section 10
Troubleshooting
Section 10 - Troubleshooting
Basic troubleshooting for Baby Pod 20 appears below. If there are questions or concerns not addressed below, take
Baby Pod 20 out of service and contact our Customer Service Department.
Problem
Baby too cold
Baby too cold
Baby too hot
Visual damage
to Baby Pod 20
Outer Shell
Buckle not fully
engaged
Baby not secure
within patient
compartment
Baby not secure
within patient
compartment
Lid opens during
transport
Possible cause
Disposable warmer mat-
tress not in use
Disposable warmer mat-
tress used in excess of two
hours
Baby Pod 20 used outside
prescribed environmental
conditions
Damaged due to sharp
objects or blunt force
Inadequate connection
Vacuum Mattress not cor-
rectly suctioned
Infant Positioning Straps
not properly secured
Baby Pod 20 Viewing Lid
not properly secured
Correction
Activate TransWarmer Mattress before placing baby
in Baby Pod 20.
Activate new TransWarmer Mattress after two hours
of transport.
DO NOT
use the TransWarmer Mattress in ambient
temperatured greater than 28.0 degrees Celcius.
Inspect inside of carbon fiber Outer Shell for cracks;
replace if necessary.
Listen for audible click to engage fully.
Suction air from Vacuum Mattress with baby in
place.
Secure Infant Positioning Straps
Check that viewing lid securing latch is properly
engaged.
Section
6.3
6.3
6.3
8
6.1
6.6
6.6
6.7