![AHT BABY POD 20 Скачать руководство пользователя страница 15](http://html1.mh-extra.com/html/aht/baby-pod-20/baby-pod-20_operating-and-maintenance-manual_2870593015.webp)
15
© 2017 • Advanced Healthcare Technology Ltd. • Baby Pod 20 ®
Section 6
Operating Instructions cont...
1.
Locate metal disc and grasp with thumb and forefingers of both hands.
2.
Flex (bend) disc rapidly until crystals begin to form.
3.
Massage pack to soften and to increase the activation rate.
4.
After pack is fully activated, indent center with palm to form a small nest.
5.
Lay infant in nest on fabric surface.
6.
Check infant skin for redness and monitor core temperature regularly.
Note: The TransWarmer Infant Transport Mattress is a single-use, disposable product. Discard after
each patient use. To order more TransWarmer Infant Transport Mattress, contact our Customer Service
Department at:
Tel: +44 (0)1787 376493
Fax: +44 (0)1787 312707
Email: [email protected]
Section 6.4 - Placing baby inside Baby Pod 20
Place baby into Baby Pod 20 making sure that the baby’s head is nearest the end marked “Head” and towards the
direction of travel denoted by the arrow on the outer shell. Incorrect positioning of the patient may cause injury
during transport.
Place baby face-up on the non-woven surface of the TransWarmer Infant Transport Mattress.
DO NOT
place a baby
directly onto the printed side of the TransWarmer mattress.
Section 6.5 - Operating Vacuum Mattress
Check to see that the Vacuum Mattress valve is at the baby’s feet. Valve should be open with the long and short
tubing sections engaged (black “O” ring hidden).
Check to see that the baby is in the desired position for transport.
Place vacuum source into short tubing section on Vacuum Mattress valve and slowly suction air from the mattress
while hand shaping the mattress around the baby. As air is removed, the Vacuum Mattress acts as a positioning
and retaining shell for the baby.
Once the Vacuum Mattress is in the desired position, remove the vacuum source and close the mattress valve by
depressing the grey tab and disengaging the long and short tubing sections. Once disengaged, the short tube’s
black “O” ring is visible.
Section 6.3 - Activating the TransWarmer Infant Transport Mattress
Warning
Patients may be harmed by the incorrect use of this device.
Warning
Improper use of the Vacuum Mattress may cause injury to the patient during transport.
Warning
DO NOT use the TransWarmer Infant Transport Mattress in ambient temperatures greater than 28.0 degrees celcius.
Ambient temperatures greater than 28.0 degrees Celcius may cause Warmer Mattress to become
too hot and cause injury to the infant.