13. APPENDIX
13.1 Technical data
Measuring input :
1 miniature thermal socket for thermocouples with
miniature thermal flat connector
A/D converter :
Delta-sigma 15-bit, at measuring rate 2.5 mops
Gain :
40, 80
Common mode input range :
-0.26 ... +2.6 V, Overload max. -4..+5V
Input current :
< 2 nA
System accuracy :
± 0.1 % v. Measuring values ± 3 Digit
Temperature drift :
0.01 %/°C
Cold junction compensation:
-30 ... +100 °C, Accuracy : ± 0.2 K ± 0.01K/°C
Self-calibration:
Automatic zero point correction
Control functions:
automatic sensor failure detection
Measuring ranges
NiCr-Ni
Type K
-200.0...+1370.0
°C
NiCroSil-NiSil
Type N
-200.0...+1300.0
°C
Fe-CuNi
Type L
-200.0... +900.0
°C
Fe-CuNi
Type J
-200.0... +950.0
°C
Cu-CuNi
Type U
-200.0... +600.0
°C
Cu-CuNi
Type T
-200.0... +400.0
°C
PtRh10-Pt
Type S
0.0...+1760.0
°C
Millivolt
-10.000...+55.000
mV
Standard equipment
LCD :
Measured value : 5x 7-segment 15 mm,
2x 16-segment 9 mm
Function 4½ x 7-segment 9 mm, 5 symbols
Operation :
7 silicone keys
Memory :
1 measuring value on the RAM
Microprocessor :
M16C28
Power supply :
Batteries :
3 AA alkaline batteries
Current consumption :
approx. 10 mA
Battery operating time :
approx. 250 hours
Housing :
(LxWxH) 127 x 83 x 42 mm ABS (acrylonitrile butadi-
ene styrene), weight : approx. 260 g
Suitable conditions
Operating temperature
-10 to +60 °C
(storage temperature
-20 to +60 °C)
Ambient relative humidity :
10 to 90 % rH
(non-condensing)
13.2 Product overview
Temperature measuring instrument THERM 2420
Order no.
1 measuring input , 2-row LCD, 7 keys, 3 AA batteries
MT 24201L
14 THERM 2420