Device configuration
10. DEVICE CONFIGURATION
On the THERM 2420 measuring instrument a
number of parameters can be configured. To
do so when switching ON press and hold down
key
MEM
. The function field should then show
an abbreviation for the parameter and the main
field should show the value currently set.
To select from all possible parameters press
keys :
M▲
or
M▼
Measuring range
Automatic switch OFF time in minutes: see 10.2
Function locking : see 10.3
To enter a value
first press :
ON
and the value starts
flashing.
To modify the value press keys :
M▲
or
M▼
Entry is completed by again pressing key :
ON
To terminate configuration
at any time press key :
MEM
10.1 Measuring ranges
The following measuring ranges can be set via menu item ´
type
´.
Transducer
Type
Measuring range Units Abbrevi-
ation
NiCr-Ni
K
-200.0...+1370.0
°C
NiCr
NiCroSil-NiSil
N
-200.0...+1300.0
°C
NiSil
Fe-CuNi
L
-200.0... +900.0
°C
FECO
Fe-CuNi
J
-200.0... +950.0
°C
IrCo
Cu-CuNi
U
-200.0... +600.0
°C
CUCO
Cu-CuNi
T
-200.0... +400.0
°C
CoCo
PtRh10-Pt
S
0.0...+1760.0
°C
Pt10
10.2 Automatic switch OFF
In menu item ´
AOFF
´ an automatic device switch OFF time can be programmed
in minutes; this will help save the batteries. This automatic device switch OFF
will not take effect if the setting is ´- -´.
THERM 2420
11
AOFF
type
Loc