11. Programming With Programming-Menus
55
measurements during a read-out to specific measurement locations or measur-
ing points (s. manual 6.7).
After selecting the function
Number
the 6-digit number is entered in the usual
way (s. 9.5). In addition to the figures 0 to 9 the characters A,F,N,P,- or _ (space)
can be used. After the input the number is activated and next to it an ´
A
´ will be
indicated until the next cyclic or manual measurement will be stored.
Function
Number:
(e.g. room 12, meas. point 1)
NUMBER:
12-001
A
Zero setting
and deactivating the number with key:
<CLR>
Activating
and
deactivating
the number with key:
<ON>
,
<OFF>
Incrementing and
activating
the number with:
< +1 >
11.2.4
Starting and Stopping of measurements
Additionally to starting and stopping of the
measurement with the keys, there are even
more possibilities, concretely conveyed by the
assistant menu
START-STOP
.
The operation via interface is described in the
manual section 6.6.
The function of start time and end time is de-
scribed here in section 11.1.4, the limit value
responses in section 11.4.3 and the varieties of
relais and trigger in section 11.6.2.
11.2.5
Scanning mode
For autonomous operation and / or for scanning by computer there are 4 scan-
ning modes available.
Normal
Internal cycle or cyclic scanning by the computer
Sleep
Internal cycle only, automatically switching off for long-term monitoring
Monitor
Internal cycle, not disturbed by computer scanning
Fail-safe
Cyclic scanning by the
PC after any failure the internal cycle
Sleepmodus
For long term monitoring with larger cycles it is possible to operate the measur-
ing device in sleep mode. Within this power saving mode the device will be
switched off after each measuring point scan (this is to be considered using
sensors with power supply!) and will be automatically switched on for the next
measuring point scan after the cycle time has expired. This procedure allows,
with one set of battery/accumulators to perform more than 15000 measuring
point scans. At 10 minutes for one cycle this results in a total measuring time of
more than 100 days.
* ALMEMO 2690-8 *
¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯
MEASURING START-STOP:
by pressing: START-STOP key
by interface comand: S2-X
by time ©
by limit values
by trigger cable
F ESC