10. Measuring using Measuring-Menus 
 

34 

10.3  

Measuring point scans and Output 

Measuring point scans can be used  to  acquire 
the measured values of all measuring points at 
particular  times,  either  manually  or  cyclically 
over a period of time, which means to store or to 
record on computer (s. man. 6.5). 
For this purpose, the menu 

 Data logger 

, e.g., is 

available: 
 

10.3.1  

Single  Output  /  Storage 

of all Measuring Points 

Single manual measuring point scans for acquir-
ing the momentary measured values of all active measuring points 

(s.man. 6.5.1.1)

 

can be triggered by the key 

 

<MANU>

 

. If the true time of day has to be indicated, 

it must first be set (s. 11.1.1). The output format can be set in the function 

 Cycle 

Timer 

 (s. 10.3.2). 

Single manual measuring point scan:

 

 

<MANU>

 

The following symbols will, 

for a short period,

 be indicated in the 

status line

The start arrow will light up and then go off again 

´

©

 ´

 

Lightening up in case of data output via interface  

´

COM 

´

 

Indicated when measured values are being stored (s. 11.1.2) 

´

REC 

´

 

Each time the key is pressed again after this, the measured values are equally 
processed with the corresponding measuring time. 

10.3.2  

Cyclic  Output  /  Storage  of  all  Measuring 

Points 

For cyclic outputs of measured values (s. man. 6.5.1.2) and recording, the cycle 
and  the  output  format  have  to  be  programmed.  The  measurement  can  be 

started

 by pressing the key 

 

<START>

 

 and 

stopped 

with the key 

 <STOP> 

. At 

each start of a measurement the max, min and average values of all measuring 
points are cleared, if the device has been correspondingly configured (standard 
setting, s. 11.5.8). 
The function 

Cycle Timer 

 indicates the cycle as long as a measurement has not 

been started. After selecting the function (s. 9.4), the cycle can be entered im-
mediately (s. 9.5). After starting a measurement you can see the timer counting 
down to the next cycle. 
Function 

 Cycle Timer 

Cycle Timer:  

00:02:00

 S

 

     Cycle (

hh:mm:ss

), Memory On, Format List 

 

The key 

 <FORM> 

 can be used as the quickest way to set the required Output 

Format (printouts s. man. 6.6.1). 

C © REC COM  l© ©l R01 * ´´´´´´µµµµ¶

 

±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±

            

                  

                                                                                                                               

 

Time: 12:34:56  Date:01.01.04 

Cycle timer:  

00:00:30 nS 

Memory free:  

508.3 kB 

Number:  

01-001  A 

01:   244.5 °C 

NiCr  Temperatur     »   H   º 

 

7 Limit max:  

250.0 °C 

Maximum value:  

245.7 °C 

7 Limit min:  

230.0 °C 

Minimum value:  

   224.1 °C 

 

START MANU  M  PRINT  ESC  

 

Содержание ALMEMO 2690-8A

Страница 1: ...____________________________ Operating Instructions Data Logger ALMEMO 2690 8A V4 7 20 10 2021 www ahlborn com...

Страница 2: ...DKU5 5V 0 4A DC Cable isolat ZA2690 UKx 10 30V 4 Sleep LED 5 LCD display Status line C Cont measuring point scan ll Measurement Start Stop REC Memory record COM Output of meas values l l Meas start en...

Страница 3: ...asurement 14 5 1 3 Process flow control 15 6 Initial Operation 17 7 Power supply 18 7 1 Rechargeable battery operation voltage monitoring 18 7 2 Battery operating mode 18 7 3 Mains operation 19 7 4 Ex...

Страница 4: ...nction Compensation 33 10 3 Measuring point scans and Output 34 10 3 1 Single Output Storage of all Measuring Points 34 10 3 2 Cyclic Output Storage of all Measuring Points 34 10 3 3 Memory space Memo...

Страница 5: ...50 11 1 2 Cycle with Storage Activation and Output Format 50 11 1 3 Conversion Rate Continuous Measuring Point Scan 51 11 1 4 Time and Date of Start Time and Date of End 52 11 2 Data Memory 53 11 2 1...

Страница 6: ...annel 1 69 11 4 7 Reference Channel 2 or Multiplexer 69 11 4 8 Element Flags 69 11 5 Device Configuration 70 11 5 1 Device Designation 70 11 5 2 Device Address and Networking 70 11 5 3 Baud Rate Data...

Страница 7: ...2 Table of Contents 7 14 1 Technical Data 81 14 2 Product overview 82 14 3 Index 83 Your contact 88...

Страница 8: ...chapter 12 Troubleshooting Should there actually be a defect please use the original packing material for depatch if possible and enclose an expressive description of the trouble in connection with th...

Страница 9: ...oard boxes protective plastic packaging materials and all preservative substances separately and in the proper manner The disposal of the device itself also of device parts accessories and con sumable...

Страница 10: ...junction with unsuitable peripheral equipment Use of the device for any purpose other than that for which it is intended Damage caused by electrostatic discharge or lightning DANGER Risk of fatal inju...

Страница 11: ...heir own integrated power supply are not electrically iso lated from one another 4 2 Handling batteries rechargeable batteries correctly When inserting batteries rechargeable batteries ensure that the...

Страница 12: ...s Centronics manual section 5 2 The entire ALMEMO networking system manual section 5 3 All functions and their control via the interface manual section 6 A complete interface command list with all pri...

Страница 13: ...o any ALMEMO meas uring instrument and are interchangeable without requiring any settings Function Channels Maximum minimum average values and differences of certain measuring junc tions can be progra...

Страница 14: ...a measuring rate of 10 mops and the data thus obtained is listed on the display However if you want to have the measured values for the se lected channel sent to an analog output it may be an advanta...

Страница 15: ...gnal the real time clock or an exceeding of limit values Time and Date The real time clock with data function or the pure measuring time are used for an accurate recording of any measurement Start and...

Страница 16: ...hem are soft keys can be used to operate the device This also allows you to fully program the sensors the device and the process control Output All data logs menu functions and stored measured values...

Страница 17: ...ingle output storage s 10 3 1 MANU or F2 Cyclic measurement Select Cycle Timer PROG Enter cycle hh mm ss s 9 5 Cycle Timer 00 05 00Sn Output format list columns n table t FORM or F3 Stop programming E...

Страница 18: ...ity drops to approx 10 the battery symbol in the status bar of the display will start to flash As and when this happens you should connect mains adapter ZB 1112 NA7 12 V 1 A and re charge the batterie...

Страница 19: ...supply voltage requirements of the sensors to 6 V 200 mA 9 V 150 mA or 12 V 100 mA s 11 4 2 see also the power supply menu 11 7 Other voltages 12 15 or 24 volts or references for a potentiometer and s...

Страница 20: ...M of the sensor connector the calibration and the programmed parameter of the device is fail safe stored in the EEPROM of the device Time and date are buffered by a lithium battery so that storage of...

Страница 21: ...has 5 input sockets 1 that the measuring channels M0 to M4 are initially allocated to However ALMEMO sensors can if required provide up to 4 channels so that 20 channels are avail able with 5 input s...

Страница 22: ...flow any compensation currents between sensors power supply and peripherals This is the case when all measuring points are on the same potential or when unequal potentials are electrically isolated T...

Страница 23: ...a a USB data supply cable ZA1919 DKU5 or a simple ALMEMO supply connector ZA1000 FSV It will also be possible via a non electrically isolated analog output cable to perform measuring operations at poi...

Страница 24: ...the selected menu with key or PROG Back to the last meas menu using the key once Back to the last programming menu using the key again Back to the menu selection with key ESC The device designation i...

Страница 25: ...ted ll or Measuring point scan started including storing REC Measuring point scan started incl output via interface COM Start end time of measurement is programmed l or l Relay status external output...

Страница 26: ...ue Clearing the programmed values CLR For programming press the key PROG Then you are in the input mode P in the middle of the softkey line the cursor flashes below the first cursor position Cycle Tim...

Страница 27: ...ion MEASURING Menus and differ in the number of measuring points 1 to 20 in the dis play of measured values in different sizes of the digits 4 8 12 mm or as bar or line chart and in the arrangement of...

Страница 28: ...ime be determined and stored from the acquired measured values of each measuring point The functions listed below are available to display these values function channels are available put them out s 1...

Страница 29: ...tions or at certain times in order to check the deviation from this reference value After selecting the func tion Meas Value s 9 4 in any menu a help window shows you all possibilities of correction o...

Страница 30: ...base value In the case of dynamic pressure probes the zero point error is al ways written temporarily i e until you switch off to the calibration off set even if the channel is locked 10 2 3 Sensor A...

Страница 31: ...o point adjustment is also possible for other sensors Additionally to the zero point adjustment 10 2 2 the slope can be corrected with a second measuring point using the function Setpoint Press ing a...

Страница 32: ...3 2 10 2 6 Atmospheric Pressure Compensation Some measuring variables depend on the environmental atmospheric pressure s 11 3 9 measuring range list w PC As a result higher deviations from the normal...

Страница 33: ...arately for each socket As and when necessary the cold junc tion temperatures can be displayed and recorded via function channel CJ see 11 3 10 These can also be used as device temperature This cold j...

Страница 34: ...ed again after this the measured values are equally processed with the corresponding measuring time 10 3 2 Cyclic Output Storage of all Measuring Points For cyclic outputs of measured values s man 6 5...

Страница 35: ...y monitor the available memory space By selecting this function two soft keys are available for a direct output and clearing of the memory The output format corresponds to the setting in the cycle s 1...

Страница 36: ...ser leaves the menu in this case do not change the meas point Limits if activated are entered as dotted lines The functions Analog Start and Analog End of the menu Special functions s 11 4 4 can be us...

Страница 37: ...can call up an own assistant menu for all modes to put in the necessary parameters and to learn the operation by help windows Measuring menu Average Value However most of the averaging functions can a...

Страница 38: ...ont 10 4 2 Averaging Mode A detailed description of averaging over measuring point scans is given in the manual section 6 7 4 The type of averaging is determined through the func tion Averaging Mode W...

Страница 39: ...2640 measurements are to be carried out in a net of individual points in a cross section which is placed vertically with respect to the flow direction s man 3 5 5 A special menu for net measurements...

Страница 40: ...average val on start s 11 5 8 Check or after selecting the average value by using CLR Starting the averaging using the key START Reading the meas time s 10 4 6 in function Meas Time 00 01 23 40 Stoppi...

Страница 41: ...n at 0 All the normal start stop functions can be used in addition the start stop output and zero setting of the timer can also be triggered by actions in the event of limit values being exceeded see...

Страница 42: ...ailable s 11 3 9 If you do not wish to program reference channels and the measuring points to be averaged begin with M0 you need simply to program the function chan nel M n to the 2nd channel of the l...

Страница 43: ...he average flow velocity can also be determined by single net measurements according to VDI VDE 2640 s 10 4 4 e g 13 24 m s To display output and or save the number of measuring operations a special f...

Страница 44: ...art the functions Analog start and Analog end in the Special functions menu should be used see 11 4 4 Having selected these functions they can be entered by pressing PROG and or directly on the approp...

Страница 45: ...ing comment MeasValList Comment 00 23 12 C Temperature Measured value including maximum value MeasValList Max value 00 23 12 C 32 67 C Measured value including minimum value MeasValList Min value 00 2...

Страница 46: ...art measuring by means of START Stop measuring by means of STOP 10 6 2 Wet Bulb Globe Temperature The stress caused in heat exposed work places can be evaluated in terms of their wet bulb globe temper...

Страница 47: ...rature ZERO ADJ o 15 Meas value medium 3 lines 00 1234 5 C ZERO ADJ o 16 Meas value large 7 lines 00 Temperature C 1234 5 ZERO ADJ o 17 Meas value bar chart 2 lines 5 0 S220 mls 15 00 o 34 Limit value...

Страница 48: ...int s 10 2 4 Setpoint 1100 0 C OFF ADJ o 45 Meas time s 10 4 6 Meas Time 00 00 00 00 CLR o 46 Meas Duration s 10 4 6 Meas duration 00 00 00 CLR o 47 End of menu o 99 10 7 2 Configuration of the menus...

Страница 49: ...Factor FACTOR 01 1 0350E 1 P07 Zero Point Correction ZERO CORR 01 0000 7 C f1 P06 Slope Correction SLOPE CORR 01 1 0013 f1 P07 Analog Start ANALOG START 01 0000 0 C P16 Analog End ANALOG END 01 0100 0...

Страница 50: ...fied format in the first line on the left side and the date on the right side s 9 5 Function Time and Date Time 12 34 56 Date 01 05 00 Format of time and date hh mm ss dd mm yy 11 1 2 Cycle with Stora...

Страница 51: ...n are displayed next to the cycle Cycle 00 15 00 Sn 11 1 3 Conversion Rate Continuous Measuring Point Scan If required for measuring point scans the conversion rate can in function Conv rate be increa...

Страница 52: ...adversely affected by interference over the connection lines wherever possible use twisted wires At 100 or 500mops saving to memory is only possible with the micro SD card but not with the internal EE...

Страница 53: ...he external memory is connected at the start of any measuring operation it will be used In the course of the measuring operation it must not be unplugged this would cause temporarily buffered measured...

Страница 54: ...Available free memory space Memory free 217 5 kB Linear memory without overwriting of data Ring memory Activate ring memory with overwriting of data ON Nb of act channels to calculate the storing time...

Страница 55: ...11 1 4 the limit value responses in section 11 4 3 and the varieties of relais and trigger in section 11 6 2 11 2 5 Scanning mode For autonomous operation and or for scanning by computer there are 4...

Страница 56: ...omputer Internal cyclic scanning is not influenced in any way by software scanning in Win Control safe initialization must be switched off The internal cycle is started as and when the software starts...

Страница 57: ...ion Number select number with keys FIRST NEXT LAST To select a time interval Enter the start time using the format hh mm ss Start time 07 00 00 Enter the end time using the format hh mm ss End time 17...

Страница 58: ...locking mode allows the rest appears in grey Output of the sensor programming of all active measuring points command P15 s man 6 2 3 with key PRINT 11 3 1 Selecting the Input Channel To query or to p...

Страница 59: ...nce for atmospheric pressure compensation N This activates conversion to standard conditions for flow measuring s 10 4 9 The remaining 8 characters can be used for the user s own descriptions at the e...

Страница 60: ...riable it is in most cases necessary to set a zero point shift and to perform a multiplication with a certain factor The functions BASE and FACTOR are avail able for this A detailed description of the...

Страница 61: ...re value in degree Celsius will be converted into degrees Fahrenheit The cold junction compensation can be switched off by using the characters C The following dimensions are automatically generated e...

Страница 62: ...EL18 Au FeCr ZA 9000 FS 270 0 60 0 C AuFe W5Re W26Re C ZA 9000 SSC 0 0 2320 0 C WR26 Ntc Type N ZA 9000 FS 30 00 125 00 C Ntc Ntc Typ N ZA 9040 SS3 0 000 45 000 C Ntc3 Ptc Typ Kty84 ZA 9040 SS4 0 0 20...

Страница 63: ...O2 Saturation with TC a PC FY A640 O2 0 260 O2 S O2 Concentration with TC FY A640 O2 0 40 0 mg l O2 C Function channels s 11 3 10 Mixture ratio wth PC FH A646 0 0 500 0 g kg H AH Dew point temperature...

Страница 64: ...nder the sub heading Function channels there is a group for representing function pa rameters in measured value processing or calculated results from linking certain measured values on measuring chann...

Страница 65: ...nse 11 3 11 Special meas ranges linearization cali bration Thanks to the new ALMEMO special connectors with extra memory for addi tional data larger EEPROM code E4 the following tasks can now for the...

Страница 66: ...corresponding measuring point will only be put out once in ten times however if 00 is programmed it will not be output The data recording also allows to suppress unnecessary measured values and as a r...

Страница 67: ...ode also has to be set as variant 2 in the output module s 11 6 2 man 6 10 9 Output socket A2 Setting the relay module to variant 2 EA Trigger Alarm relay int allocated 2 Rx int allocated Activating r...

Страница 68: ...programm all param eters of an analog output 11 4 5 Output Function If not the real measured value the measuring point Mxx but only the Max Min Average or Alarm value is required then this function c...

Страница 69: ...after Reference channel 1 In all other cases it is possible to change the pin assignment in the connector by changing the input multiplexer with the function Multiplexer s manual 6 10 2 Programming o...

Страница 70: ...networked to centrally acquire the measured values of several instruments that are located at different places see manual 5 3 For communicating with networked devices it is mandatory that all devices...

Страница 71: ...the display can be switched on or off in 3 levels in the selection menus by using the key ON or in the function Lighting of the device configuration attention the power consumption is fife times highe...

Страница 72: ...ction temperature socket temperature CJ Temperature 25 4 C 11 6 Output Modules The data logger ALMEMO 2690 8 has two output sockets A1 and A2 to allow an output of the measured values as analog or dig...

Страница 73: ...rigger analog adapter ZA 8006 RTA3 offers up to 10 relays or options with 2 of these as trigger inputs or up to 4 as analog outputs All elements can be configured individually for these functional var...

Страница 74: ...rmits manual activation of the relays via the keypad or via the interface see Manual 6 10 10 Relay variant 8 8 Driven externally For manual activation of relays press ON or OFF The watchdog setting fo...

Страница 75: ...ut in variant 0 In this mode the semi continuous measuring rate see 11 1 3 is the most suitable because in this way the analog output will be processed most frequently Assigning an analog output to a...

Страница 76: ...ime using a mains adapter 12V minimum 1 5A irrespective of their current charge status Using mains adapter ZA 1312 NAx 1 5 A charging 2000 mA batteries lasts up to 3 hours higher capacities will take...

Страница 77: ...from memory 3 recording to memory and output from memory 4 wizard menus 5 measuring menus except user menu U1 Fct Locking the functions and keys 0 none 1 data input switching on and off PROG ON OFF Z...

Страница 78: ...ent influences from disturbances by shielding or twisting Error Data transmission via interface does not function Remedy Check interface module connections and setting Are both de vices set to the sam...

Страница 79: ...ssibly control printouts Additionally the software ALMEMO Control al lows to print the monitor pages including the programming and also to save an extensive function test in the device list or the ter...

Страница 80: ...80 13 DECLARATION OF CONFORMITY...

Страница 81: ...time clock buffered with Lithium battery Microprocessor M16C62P Power supply Internal with recharg batteries 3 rech batteries AA Eneloop type low self discharge LSD NiMH 2 Ah Integrated battery charg...

Страница 82: ...OA 2690 KL Accessories Memory connector including SD card minimum 128 MB RS ZA1904 SD Mains adapter with ALMEMO connector 12V 1 5 A ZA1312 NAx Accumulator set with 3 NiMH Mignon cells 1600 mAh coded...

Страница 83: ...nnection of the Transducers 21 Continuous Measuring Point Scan 51 Contrast 71 Control Symbols 25 Conversion Rate 51 Correction of the Meas Value 29 Correction Values 61 Current output 75 Cycles 50 Cyc...

Страница 84: ...ctor 66 Process flow control 15 Product overview 81 Programmed analog value output 75 Programming Menus 24 50 Rechargeable battery 18 Reference Channel 1 69 Reference Channel 2 69 Reinitialisation 19...

Страница 85: ...uts 74 Trigger modules 73 Troubleshooting 78 Two point Adjustment 31 U Sensor Min 66 User Menus 47 Volume Flow Measurement 43 Waste disposal 9 Wet Bulb Globe Temperature 46 Zero Point Adjustment 30 Ze...

Страница 86: ...Notes 86...

Страница 87: ...Notes 87...

Страница 88: ...Mess und Regelungstechnik GmbH Eichenfeldstra e 1 83607 Holzkirchen Germany internet http www ahlborn com e mail amr ahlborn com Despite great care incorrect information cannot be ruled out Technical...

Отзывы: