17 Technical Information
Window Meanings
244/288
Turbo-V 1001 Rack Controller User Manual / 87-901-033-01
Soft Start
“Soft Start” mode is provided to start the pump after a protracted
stop of the Turbo 1001 Pump. This allows a better grease distribution
in the bearings.
The “Soft Start” mode is disabled by default. The “Soft Start” mode
can be activated by a suitable software, by Remote I/O or by Front
Panel command (see the paragraphs “INTERCONNECTIONS” and “RS
232/485 COMMUNICATION DESCRIPTION”). “Soft Start” frequency
steps are the followings: 150 Hz, 250 Hz, 350 Hz, 450 Hz, 550 Hz and
650 Hz. The pump remains at each step for a period of 300 s. A fail
condition occurs if the pump is not able to properly spin up the
driving frequency (Too High Load Error).
Содержание X3501-64003
Страница 4: ...Turbo V 1001 Rack Controller 4 288 Turbo V 1001 Rack Controller User Manual 87 901 033 01 ...
Страница 282: ...17 Technical Information Orderable Parts 282 288 Turbo V 1001 Rack Controller User Manual 87 901 033 01 ...
Страница 283: ......
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......