Running the Calibration Wizard
7
Agilent SPS 4 Autosampler User’s Guide
79
10
When all four points have been captured, a confirmation message is
displayed asking if you want to write the new calibration values.
•
Click
Yes
to write the calibration values.
The autosampler is adjusted using the new calibration value.
•
Click
No
if you want to retry.
The calibration process is restarted.
11
After calibration has been successfully completed, check the positioning
of the sample probe by clicking on each corner position button, and
driving the Z- Axis position down. Confirm that the sample probe is
positioned in the center of each well.
Error List
On running the calibration wizard software, you may encounter any error
listed below:
• Serial Port Error
Example: Autosampler did not respond.
A serial port error occurs if the software cannot communicate with the
autosampler. Check that the power is on, and no error lights are
illuminated. Also confirm the USB connection and the COM port setting
in the instrument control software. If the problem still remains
unresolved, contact your Agilent customer support representative.
Содержание SPS 4
Страница 1: ...Agilent Technologies Agilent SPS 4 Autosampler User s Guide...
Страница 28: ...28 Agilent SPS 4 Autosampler User s Guide 2 Introduction This page is intentionally left blank...
Страница 52: ...52 Agilent SPS 4 Autosampler User s Guide 3 Installation This page is intentionally left blank...
Страница 62: ...62 Agilent SPS 4 Autosampler User s Guide 5 Maintenance This page is intentionally left blank...
Страница 96: ...96 Agilent SPS 4 Autosampler User s Guide A Appendices This page is intentionally left blank...
Страница 97: ......