74
Agilent SPS 4 Autosampler User’s Guide
7
Running the Calibration Wizard
Performing User Calibration
To perform user calibration:
1
Ensure the autosampler power is on.
The sample racks fitted should be empty and not contain any sample
vials. If a microtiter plate Adapter is fitted, it should be loaded with
four 96 well microtiter plates.
2
Start the Calibration Wizard software. The Calibration Wizard window
opens as shown below.
3
Select the appropriate COM port from the drop- down list, and then
click
Connect
in the
Com Port
field.
If you chose the wrong one, an error message appears. If the
appropriate COM port number is not listed, check the autosampler
power, the USB connection, and the COM port setting in the instrument
control software.
4
If the calibration wizard can connect to your autosampler successfully,
the field under the
Com Port
field shows the information on the rack
layout specific to your autosampler. Ensure that the sample rack types
displayed on the screen correspond to those loaded into their respective
positions on the rack location mat.
N O T E
Close the instrument software before launching the Calibration Wizard.
Содержание SPS 4
Страница 1: ...Agilent Technologies Agilent SPS 4 Autosampler User s Guide...
Страница 28: ...28 Agilent SPS 4 Autosampler User s Guide 2 Introduction This page is intentionally left blank...
Страница 52: ...52 Agilent SPS 4 Autosampler User s Guide 3 Installation This page is intentionally left blank...
Страница 62: ...62 Agilent SPS 4 Autosampler User s Guide 5 Maintenance This page is intentionally left blank...
Страница 96: ...96 Agilent SPS 4 Autosampler User s Guide A Appendices This page is intentionally left blank...
Страница 97: ......