Agilent SPS 4 Autosampler User’s Guide
7
Running the Calibration Wizard
Before User Calibration
Check the following points before performing user calibration. Performing
user calibration when the unit and accessories have not been setup
correctly could cause a hazard.
•
The spill tray is correctly fitted.
•
The rack location mat is seated correctly in the spill tray and end
frames.
•
The sample racks are correctly engaged in the locating holes in the rack
location mat and are seated flat on the rack location mat.
•
A sample probe is fitted and has been mounted correctly.
•
The sample probe is essentially straight.
Calibration Wizard Overview
Use the calibration wizard to adjust the centering of the sample probe.
There are four calibration points, one for each corner in the movement
range of the probe arm. At each calibration point, adjust the probe
position.
The following shows the Calibration Wizard window.
Figure 19
Calibration Wizard Window