RS-232 and RS-485 Options for the senTorr Gauge Controller
1-4
DRA
FT
❑
RS–232 Standard
Table 1-4 lists the serial port pin-outs of the standard 9-pin connector for both
standard and null modem operation. Table 1-5 lists DIP switch SW1 settings to select
the signal output mode. The factory setting is for the null modem mode.
Table 1-4
RS–232 Signal Communications
Signal
Null Modem Pin
Standard Pin
Description
Gnd
5
5
Signal Ground
TXD
3
2
Transmitted Data
RXD
2
3
Received Data
RTS
8
7
Request To Send
CTS
7
8
Clear To Send
DTR
4
4
Data Terminal Ready
Table 1-5
RS–232 Null Modem or Standard Selection
Switch Number
Null Modem Setting
Standard Setting
SW1-1
closed
open
SW1-2
open
closed
SW1-3
closed
open
SW1-4
open
closed
SW1-5
closed
open
SW1-6
open
closed
SW1-7
closed
open
SW1-8
open
closed