RS-232 and RS-485 Options for the senTorr Gauge Controller
DRA
FT
1-3
4. Set the DIP switches as per board type, as discussed below:
❑
RS–485
Refer to Table 1-2. If the senTorr is shipped with this option installed, the board is
configured as ordered.
A differential terminating resistance of 220 Ohms can be switched into the two-wire
pairs. The factory setting is unterminated (Table 1-3).
Table 1-2
RS–485 Selection
Switch Number
RS–485
SW1-1
closed
SW1-2
open
SW1-7
closed
SW1-8
open
Table 1-3
RS–485 Terminating Resistance Selection
Switch Number
Number Terminating
Resistance
220 Ohm Terminating
Resistance
SW1-3
open
closed
SW1-4
open
closed
Содержание RS-232
Страница 6: ...RS 232 and RS 485 Options for the senTorr Gauge Controller This page intentionally left blank ...
Страница 8: ...RS 232 and RS 485 Options for the senTorr Gauge Controller This page intentionally left blank ...
Страница 10: ...RS 232 and RS 485 Options for the senTorr Gauge Controller This page intentionally left blank ...
Страница 14: ...RS 232 and RS 485 Options for the senTorr Gauge Controller This page intentionally left blank D R A F T ...
Страница 27: ......