RS-232 and RS-485 Options for the senTorr Gauge Controller
DRA
FT
1-9
Reset senTorr
#aa06t
>t
Read Remote Input State
#aaODt
>OOhht
where 00 = inactive (low)
01 = active (high)
Read Pressure Dump
#aaOFt
>x.xxxE-xx[,. ..] t
The number of bytes in the
response varies with the
configuration of the unit. The
order of readings is from top to
bottom of the front panel
display.
Set pressure units to Torr
#aa10t
>t
Set pressure units to mBar
#aa11t
>t
Read pressure units
#aa13t
>hht
where:
❑
hh = 00 is Torr
❑
hh = 01 is mBar
Set key pad lock OFF
#aa20t
>t
Set keypad lock ON
#aa21t
>t
Read keypad lock status
#aa22t
>hht
where:
❑
hh = 00 is unlocked
❑
hh = 01 is locked
❑
hh = is a partial lock
Set Partial keypad lock
#aa23t
>t
Set Emission OFF
#aa30lnt
>t
Set Emission ON
#aa31lnt
>t
Read Emission status
#aa32lnt
>hht
where:
❑
hh = 00 is OFF
❑
hh = 01 is ON
Set degas OFF
#aa40l1t
>t
Set degas ON
#aa41l1t
>t
Table 1-8
Serial Command Set (Continued)
Function
Command
Response