Plumbing the Purge Gas
QuickSwap Installation and Startup
9
6
Remove the valve box cutout using a side cutter (
7
Install the 6890 valve box blank provided in the QuickSwap supplies and
spares kit using the two screws. Note that the notch in the blank should be
on the left and toward the back.
8
Locate the tubing from channel 5 of the EPC module. Unroll the tubing
through the slot in front of the EPC module, then along the back of the oven
top toward the hole in the valve box blank.
WA R N I N G
The insulation around the oven is made of refractory ceramic fibers. To avoid inhaling
fiber particles, we recommend the following safety procedures: ventilate your work
area; wear long sleeves, gloves, safety glasses, and a disposable dust/mist respirator;
dispose of insulation in a sealed plastic bag; wash your hands with mild soap and cold
water after handling the insulation.
Figure 2
Remove the valve box cutout
Cut tabs around
edges and
remove
C A U T I O N
Avoid kinks or sharp bends in the tubing.