![Agilent Technologies QuickSwap G3185B Скачать руководство пользователя страница 33](http://html1.mh-extra.com/html/agilent-technologies/quickswap-g3185b/quickswap-g3185b_installation-and-setup_2867789033.webp)
Overview
QuickSwap Installation and Startup
33
3
Select
MoreParams
, then
Tune Params
to display the
Tune & Display Parameters
screen (
4
Enter the target
Masses
in the
Mass 1
,
Mass 2
, and
Mass 3
fields. Use
m/z
28
and 32 to monitor air and
m/z
83 for a fragment ion of
Freon 134a
.
5
Enter a
Scan Range
of
25.00
to
125.00
.
6
Select
OK
or click in any other field to enter new values.
7
Select
Scan
to begin a continuous scan at these masses.
The scan in
shows air background immediately after pumpdown.
Examine the
m/z
28 and 32 peaks first. If they are below 20,000 to 30,000,
proceed to Step
. If not, check for gross leaks in the analyzer section of the
MSD (for example, whether the side door is closed fully or the vent valve is
open). If high 28
m/z
is seen, tighten QuickSwap onto the transfer line until
the peaks drop significantly.
Figure 14
Tune & Display Parameters