6
Installation Note
E4440-90632
Installation Procedure
Installation Procedure
For assistance at any time during this procedure, get in touch with your nearest Agilent
Technologies Sales and Service Office. To find your local Agilent office access the
following URL:
http://www.agilent.com/find/assist
Preliminary Hardware Verification
1. Check for the presence of Option 1DS by pressing
System
,
Show System
, and look at
the option list.
• If Option 1DS is present it must be uninstalled before Option 110 can be installed.
a. Delete Option 1DS License by pressing
System
,
More
,
More
,
Licensing
,
Option
.
Use the front panel numerical key pad to enter the option designator 1DS and
press the
Enter
key. Press
Delete Option
.
b. Check that Option 1DS has been removed by pressing
System
,
More
,
Show
System
, and looking at the option list.
c. Follow the instructions in this installation note in the section named
to remove the Option 1DS hardware.*
• If Option 1DS is not present then continue with the installation of Option 110.
*Option 110 and Option 1DS are not compatible. Damage to the instrument may
result if they are both installed.
2. Check for the presence of Option 123, Preselector Bypass by pressing
System
,
More
,
Show System
, and look at the option list.
• If Option 123 is present then follow the instructions in this installation note in the
“Installing the Preamplifier with Option 123”
• If Option 123 is not present then follow the instructions in this installation note in
the section named