(Optional) Step 7. Verify Operation > 3 GHz
Operation verification is a test that, when completed, will ensure that the
N8201A performance downconverter is operating correctly in the high band (> 3 GHz).
Performing a Self-Test
Downconverter to Spectrum Analyzer Connections
•
Downconverter: 7.5 MHz Out
•
Spectrum Analyzer: RF Input
Downconverter to Source Connections
•
Downconverter: RF Input
•
Source: RF Output
Perform the following procedure to run a self-test:
1
From the Windows Desktop,
select
Start
>
All Programs
>
Agilent SI Tools
>
Agilent Synthetic Instrument GUI
.
C A U T I O N
Make sure that the total power of all signals at the
N8201A performance
downconverter
input
does not exceed
+30 dBm (1 watt).