84
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz
4
Front and Rear Panel Features
1588 LED
The IEEE 1588 LED clock status has the following states:
COHERENT CARRIERS
3-7 GHz 1st LO In
Provides the 1st LO input signal (3 to 7 GHz) to the downconverters
internal LO distribution circuitry.
3-7 GHz 1st LO Out
Provides the 1st LO output signal (3-7 GHz at +15 dBm max) as a
coherent carrier reference for phase locking two receivers.
3.6 GHz 2nd LO In
Provides the 2st LO input signal (3.6 GHz) to the downconverters
internal LO distribution circuitry.
3.6 GHz 2nd LO Out
Provides the 2nd LO output signal (3.6 GHz at +3 dBm) as a
coherent carrier reference for phase locking two receivers.
State of the Clock
Clock Status
Illumination
OFF
Not synchronized
None
ON
Synchronized, clock is IEEE 1588 Slave
Solid green
ON
Synchronized, clock is IEEE 1588 Master Blinking green (once
every second)
ON
Synchronized, clock is IEEE 1588 to
Grand Master
Blinking green (once
every two seconds
OFF
IEEE 1588 is in a fault state
Solid red
N O T E
Do not remove the cables connecting the coherent carrier inputs and outputs. These cables
are connected at the factory and must remain so for proper instrument operation.