Service Guide N5230-90014
4-15
PNA Series Microwave Network Analyzers
Troubleshooting
N5230A
Power Up Troubleshooting
Verifying the A3 Front Panel Interface Board
To verify that the HSYNC (horizontal sync), VSYNC (vertical sync), and LCD clock are
functioning correctly, measure the signals listed in
. If all of these signals measure correctly, suspect a defective backlight or LCD.
The backlight is the most probable cause.
CAUTION
Be careful not to short connector pins together when measuring these signals.
Figure 4-6 Verifying HSYNC, VSYNC, and LCD Clock
If any of the three signal types is incorrect, replace the A3 front panel interface board.
Refer to
“Removing the A3 Front Panel Interface Board” on page 7-10
.
If all of the signal types are correct, replace the A2 display assembly. Refer to
the A2 Display Assembly” on page 7-10
Table 4-4
A3 Front Panel Interface Board, Voltages and Signals
Signal Type
Test Point
Voltage
Signal
HSYNC
J2 pin 3
0 to +3 V
30.8 kHz square wave
VSYNC
J2 pin 4
0 to +3 V
60 Hz square wave
CLOCK
J2 pin 27
0 to +3.4 V pk-to-pk
25 MHz sine wave
Содержание N5230A
Страница 10: ...Contents 6 Service Guide N5230 90014 Contents...
Страница 11: ...Service Guide N5230 90014 1 1 1 Safety and Regulatory Information...
Страница 19: ...Service Guide N5230 90014 2 1 2 General Product Information...
Страница 31: ...Service Guide N5230 90014 3 1 3 Tests and Adjustments...
Страница 71: ...Service Guide N5230 90014 4 1 4 Troubleshooting...
Страница 121: ...Service Guide N5230 90014 5 1 5 Theory of Operation...
Страница 147: ...Service Guide N5230 90014 6 1 6 Replaceable Parts...
Страница 183: ...Service Guide N5230 90014 7 1 7 Repair and Replacement Procedures...
Страница 229: ...Service Guide N5230 90014 8 1 8 General Purpose Maintenance Procedures...
Страница 262: ...Index 14 Service Guide N5230 90014 Index...