3-14
Service Guide N5230-90014
Tests and Adjustments
PNA Series Microwave Network Analyzers
Preliminary Checks
N5230A
Cable Insertion Loss Check
1. With the test port cable still connected to Port 1, connect a short to the other end of the
cable.
2. From the
Marker
menu, click
Marker Search
. In the
Marker Search
dialog box, in the
Search Type
box, select
Minimum
. Click
Execute
, and then click
OK
.
3. The displayed response is twice the actual loss. To get the actual worst case insertion
loss, divide the value at the marker annotation by two. Refer to the cable manual to see
if it meets the insertion loss specification. For an example of a typical insertion loss
measurement, see
.
Figure 3-5
Typical Cable Insertion Loss Response
If the Cable Insertion Loss Check Fails
1. Clean the cable and devices and torque to specification. Repeat the check.
2. If the check still fails, the cable should be repaired or replaced.
Содержание N5230A
Страница 10: ...Contents 6 Service Guide N5230 90014 Contents...
Страница 11: ...Service Guide N5230 90014 1 1 1 Safety and Regulatory Information...
Страница 19: ...Service Guide N5230 90014 2 1 2 General Product Information...
Страница 31: ...Service Guide N5230 90014 3 1 3 Tests and Adjustments...
Страница 71: ...Service Guide N5230 90014 4 1 4 Troubleshooting...
Страница 121: ...Service Guide N5230 90014 5 1 5 Theory of Operation...
Страница 147: ...Service Guide N5230 90014 6 1 6 Replaceable Parts...
Страница 183: ...Service Guide N5230 90014 7 1 7 Repair and Replacement Procedures...
Страница 229: ...Service Guide N5230 90014 8 1 8 General Purpose Maintenance Procedures...
Страница 262: ...Index 14 Service Guide N5230 90014 Index...