3-39
PNA Series Microwave Network Analyzers
Tests and Adjustments
N5230A
Adjustments
Figure 3-15 Setup 1 for the Receiver Calibration Adjustment
2. On the
System
menu, point to
Service
,
Adjustments
, and then click
Receiver Calibration
.
3. Ensure the GPIB settings are correct.
4. Click
Calibrate
, and then follow the instructions as they are displayed.
Figure 3-16 Setup 2 for the Receiver Calibration Adjustment