Service Guide N5230-90014
5-15
PNA Series Microwave Network Analyzers
Theory of Operation
N5230A
Signal Separation Group Operation
Signal Separation Group Operation
The signal separation group divides the source incident signal into a reference path and a
test path. Refer to
• The reference signal are transmitted to the receiver group as the R1 and R2 inputs.
• The test signal is transmitted through—and reflected from—the device under test
(DUT) and then is transmitted to the receiver group as the A and B inputs.
• Control lines to this group are routed from the A16 test set motherboard.
In this section, the following assemblies are described:
•
A19 Switch/Splitter/Leveler/Amplifier/Multiplier (SSLAM)
•
A21 and A22 Test Port Couplers
•
A25 and A26 60-dB Source Step Attenuators (Option 1E1)
• Front Panel Jumpers—
Configurable Test Set (Option 014)
Figure 5-3 Passive Measurement Configuration