Operation
Clock and Data Demultiplexer 44 Gb/s User Guide
19
Clean the connectors as described in the following procedure. Cleaning
connectors with alcohol shall only be done with the instruments power cord
removed, and in a well-ventilated area. Allow all residual alcohol moisture to
evaporate, and the fumes to dissipate prior to energizing the instrument.
1.
Remove any dust or loose particles using a low-pressure air source.
2.
Moisten a lint-free swab with isopropyl alcohol. Do not saturate the
swab.
3.
Minimize the wicking of the alcohol into the connector structure.
4.
Clean the mating plane surfaces and threads.
5.
Allow alcohol to evaporate, and then use a low-pressure air source to
blow surfaces clean.
6.
Make sure no particles or residue remains.
7.
Inspect connector for damage.