Operation
24
Clock and Data Demultiplexer 44 Gb/s User Guide
The clock input to the Demux is routed through the phase shifter to align the
clock edge to the data. By turning the phase shifter knob the clock edge can
be aligned to the middle of the opening in the input data eye to get an error
free output from the Demux. The phase shifter is a 60° per GHz, 18 turns
phase shifter.
The demuxed data outputs are routed thru Demux Data Out 1 through Demux
Data Out 4 connectors on the front panel. These are 2.92 mm female
connectors. They are single ended AC coupled outputs.
This Manual:http://www.manuallib.com/agilent/n4968a-demultiplexer-user-guide.html