1-22
Installing the Test Set
Step 5. Make the Interconnections between the Test Set and the PNA
NOTE
* The Reference channel path through the test set is used to compensate for
electrically long DUTs. This also results in added loss in the Reference
channel and higher trace noise.
Therefore, unless making high speed measurements of electrically long
devices, do NOT remove the PNA
REFERENCE 1
and
REFERENCE 2
front-panel loops, and do NOT use the Reference channel interconnect
cables (items 1, 2, 3, 4, 13, and 14).
For more information, see ‘Electrically long devices’ in the PNA Help
system.
NOTE:
Cables are orderable only as cable set AD00756.
Call Out
Sequence
Cable Part
Number
From
Network Analyzer
To
Test Set
1*
AD00756-1
REF 1 SOURCE OUT
REF 1 SOURCE OUT
2*
AD00756-1
REF 1 RCVR R1 IN
REF 1 RCVR R1 IN
3*
AD00756-1
REF 2 RCVR R2 IN
REF 2 RCVR R2 IN
4*
AD00756-1
REF 2 SOURCE OUT
REF 2 SOURCE OUT
5
AD00756-2
PORT 1 SOURCE OUT
PORT 1 SOURCE OUT
6
AD00756-2
PORT 1 CPLR THRU
PORT 1 CPLR THRU
7
AD00756-2
PORT 2 CPLR THRU
PORT 2 CPLR THRU
8
AD00756-2
PORT 2 SOURCE OUT
PORT 2 SOURCE OUT
9
AD00756-3
PORT 1 CPLR ARM
PORT 1 CPLR ARM
10
AD00756-3
PORT 1 RCVR A IN
PORT 1 RCVR A IN
11
AD00756-3
PORT 2 RCVR B IN
PORT 2 RCVR B IN
12
AD00756-3
PORT 2 CPLR ARM
PORT 2 CPLR ARM
13*
AD00756-4
REF 1 on rear panel of the test set
14*
AD00756-4
REF 2 on rear panel of the test set