Observation 

Suggestion 

Signal fluctuates during run 

Check lamp, replace if necessary. See 

Changing the Lamp

.

 

Verify that the raw (multicomponent) fluorescence signal is below the 
recommended maximum of 35000 and above the recommended 
minimum of 3000 RFU. If necessary, adjust the 

Filter Gain Settings

 to 

bring the signal into the recommended range. 
If ROX is being used as a reference dye, the final ROX concentration in the 
reaction should be approximately 30 nM. If the dye SYBR

®

 Green I is used 

as the reporter, ensure that the dye is diluted appropriately before addition 
to the reaction mix. 
Verify that the instrument is not subjected to vibrations from an external 
source during the run, such as from another instrument on the same 
bench. Also verify that the line power provided to the instrument is within 
the required specifications of 100–120 volts AC (+/–10%), 50/60 Hz,  
10 A for a 110V system or 200–240 volts AC (+/–10%), 50/60 Hz, 5 A 
for a 230V system. If consistent power cannot be maintained by the local 
power supply, it may be necessary to use a Line Conditioner. An example 
Line Conditioner that is known to work well is the POWERVAR ABC700-
11MED. 
Ensure the instrument is not located near a window where the back of the 
instrument is exposed to direct sunlight. If this is the case, the window 
should be blocked or the instrument should be relocated. 

Unexpected results in one sample 

Check the sample container for contaminating material.  
Check optical clarity of sample container cap. 
Ensure the liquid inside that sample tube did not evaporate during the run, 
which would indicate a vapor leak in that particular sample. Vapor leaks 
may occur if the sample was not sealed properly or incorrect plasticware 
was used. 
Clean the thermal block as outlined in the section 

Maintaining the 

Instrument

Decreased volume in sample containers at 
end of run 

Sample containers are not vapor tight. Ensure caps are tightly sealed and 
containers are not malformed. Also ensure that proper plasticware was 
used. 

Excess condensation in reaction tubes 

When using single tubes or strip tubes, install the Perfect Fit frame on the 
thermal block to optimize contact between the sample tubes and the 
heated lid. 

No amplification plots are visible in the 
Results screen 

The wrong data collection point may have been assigned to the 
amplification plots (e.g. the dissociation curve collection point). Reset the 
data collection point for the amplification plots on the 

Analysis 

Selection/Setup

 screen.

 

Ensure that a data collection point was specified on the 

Thermal Profile 

Setup

 screen. Without a data collection point, no data is collected and 

analysis is impossible. 
If you selected dRn data when a reference dye was not defined for the 
experiment, you will not see any amplification plots on the Results screen. 
In this case, there will normally be an error message under the plot stating 
“None of the wells selected have a reference dye selected.” Change the 
Fluorescence to dR and the amplification plots will be visible. 

 

Mx3000P and Mx3005P QPCR Systems Setup and User’s Guide 

31 

Содержание MX3000P

Страница 1: ...Mx3000P and Mx3005P QPCR Systems Setup and User s Guide...

Страница 2: ...Mx3000P and Mx3005P QPCR Systems SETUP AND USER S GUIDE IN 70225 J 1 For Research Use Only Not for Use in Diagnostic Procedures...

Страница 3: ...postage or shipping costs prepaid to Agilent Loss or damage during shipment shall be at the risk of the purchaser This limited warranty does not include repair or replacement necessitated by accident...

Страница 4: ...Run 15 Hardware Information and Specifications 20 Hardware Specifications 20 Cautions and Warnings Pr cautions et Avertissements 20 Instrument Status Indicator LED s 22 Sample Loading 23 Thermal Block...

Страница 5: ......

Страница 6: ...non skirted PCR plates catalog 410088 5 Mx3000P Mx3005P optical strip caps catalog 401425 60 strips of 8 caps MxPro QPCR software installation CD 1 Installation qualification test plateb and plate set...

Страница 7: ...The instrument should be installed in a location where it will not be exposed to direct sunlight or vibration during use Environmental Conditions for Use Maximum altitude 2000 m 6562 ft Ambient temper...

Страница 8: ...the instrument door by pushing up on the door handle Remove and retain the foam used to immobilize the scanning head on the track 5 Move the scan arm to the left of the thermal block top 6 To access t...

Страница 9: ...proper instrument operation 1 Remove the label located over the power entry module on the back of the instrument just above the power switch 2 Use a small flathead screwdriver or similar tool to open...

Страница 10: ...the desired voltage is in the proper orientation If your instrument is 100 120 volts insert the fuse holder such that the 115V label is right side up If your instrument is 200 240 volts insert the fus...

Страница 11: ...ded serial cable to the RS 232 serial port at the rear of the instrument If your PC has a serial port connect the free end of the cable to the serial port on the back or side of the computer If your P...

Страница 12: ...n the back of the Microsoft Office installation CD sleeve and activation of the software either through an encrypted transfer over the Internet or by telephone Activation is required for exporting the...

Страница 13: ...will guide you through the MxPro QPCR software setup process Click Next to continue with installation 7 A window will open with a copy of the MxPro QPCR software license agreement If you accept the te...

Страница 14: ...em and then click Next 9 When asked to continue in the following dialog box click Yes Important If a NET Framework Initialization Error dialog box appears please contact Technical Services for assista...

Страница 15: ...FR Part 11 Features in the installation window below A sales order number indicating purchase of these features is required to obtain an activation key If you do not require 21 CFR Part 11 compatible...

Страница 16: ...activation The activation key can be obtained by contacting the technical services department by phone E mail or via the Agilent website The system generated ID sales order number instrument serial n...

Страница 17: ...ormation Technology department Click Next once a database type is selected 14 Depending on the database option selected one of the following windows will appear Complete the appropriate fields either...

Страница 18: ...er location using the Browse button Click Next once a folder location is selected 16 Once the MxPro QPCR software files and Microsoft SQL database software files if applicable are installed to the des...

Страница 19: ...e application is opened In order to start the application enter the initial case sensitive default username Admin and case sensitive default password password Click Login 20 The first time you log in...

Страница 20: ...inutes leaving the plate sealed in the foil wrapper while thawing After thawing remove the foil wrapper and then mix the contents of the wells by completing the following actions five times Invert the...

Страница 21: ...ime that the MxPro QPCR software is opened using an instrument connected computer the Instrumentation Qualification Test Start dialog box will appear This dialog box may also be opened by selecting Qu...

Страница 22: ...e instrument is running will abort the run If the run is accidentally aborted by opening the instrument door the following error message will appear Do you want to abort or continue the run Close the...

Страница 23: ...Microsoft Excel software Validation template will open to the FAM analysis tab First verify that the Data Count cell reports the result Good ensuring that all of the data has been correctly imported...

Страница 24: ...gilent com Include a contact name the name of your institution and the instrument serial number when sending the data to Technical Services See Mx3000P and Mx3005P QPCR Systems Support for additional...

Страница 25: ...Mx3005P system 5 dyes per well Sample container configurations 200 l tube format configured as 96 well plates strip tubes or single tubes Reaction volume per tube 10 50 l Electrical power input 100 12...

Страница 26: ...appear on screen if the door has been opened and the run has been interrupted Autres Avertissements Soyez prudents lorsque vous chargez et d chargez les chantillons du bloc thermique car ce dernier po...

Страница 27: ...table below Appearance of Ready Status Indicator LED Instrument Status Continuously lit The instrument control software is operational and the unit is ready to run an experiment Blinking The unit is n...

Страница 28: ...he door should not be opened Open the door located at the front of the instrument by sliding it all the way to the top To expose the thermal block lift the heated lid by pulling forward on the handle...

Страница 29: ...he thermal block Important When using individual tubes or strip tubes as sample containers insert the Perfect Fit frame on the thermal block platform before loading the samples See Use of the Perfect...

Страница 30: ...e will be positioned in the spaces between wells on the perimeter of the thermal block surface The Perfect Fit frame should be seated on the thermal block platform with limited freedom of movement The...

Страница 31: ...well working rack Agilent Catalog 410094 Apply direct downward force to each cap and then view the rack from the side to make sure all the caps are seated evenly Alternatively caps can be applied usin...

Страница 32: ...hot The lamp is located in the excitation housing on the left hand side of the instrument Before changing the lamp be sure to turn off the instrument The lamp is replaced along with the lamp housing a...

Страница 33: ...To replace the fuse disconnect the instrument from the power source open the fuse housing cover and remove the fuse holder from the power entry module Replace the fuse with the appropriate fuse type...

Страница 34: ...ess the software will offer the option to continue displaying data from the current run If the run completes before communication is reestablished the software should offer the option to download the...

Страница 35: ...fluorophore You can test the probe by performing a nuclease digestion to ensure it is unquenching as expected incubate 100 nM of probe in 25 l of 1 buffer with 10 U of DNase or S1 nuclease at room te...

Страница 36: ...material Check optical clarity of sample container cap Ensure the liquid inside that sample tube did not evaporate during the run which would indicate a vapor leak in that particular sample Vapor lea...

Страница 37: ...ification When running samples with a broad dynamic range use the adaptive baseline algorithm If amplification curves still look abnormal it may be necessary to manually set the baseline ranges Baseli...

Страница 38: ...ion or 400 bp for SYBR Green I detection Magnesium concentration is too low run a titration to optimize concentration Insufficient or non specific product Verify product formation through gel electrop...

Страница 39: ...icensor also grants to you a nonexclusive license to use the Materials solely in conjunction with your use of Software as licensed in this Agreement Modifications You may modify or change the Software...

Страница 40: ...shall Licensor s liability to you whether in contract tort including negligence or otherwise exceed 250 00 This License Agreement is effective until terminated You may terminate this License Agreemen...

Страница 41: ...may use only the medium appropriate for your single computer You may not use or install the other medium on another computer You may not loan rent lease or otherwise transfer the other medium to anoth...

Страница 42: ...ria 01 25125 6800 Benelux 02 404 92 22 Denmark 45 70 13 00 30 Finland 010 802 220 France 0810 446 446 Germany 0800 603 1000 Italy 800 012575 Netherlands 020 547 2600 Spain 901 11 68 90 Sweden 08 506 4...

Страница 43: ...com genomics Distributors For a list of worldwide distributors please visit our website www agilent com genomics This information is subject to change without notice Agilent Technologies Inc 2015 Pri...

Отзывы: