Agilent Technologies Magnis NGS Prep System Скачать руководство пользователя страница 9

Decontaminating the System

Magnis NGS Prep System Quick Reference Guide

9

Add new user

accounts

Each user who uses the system needs an account. 

1

From the Home screen, press 

Settings

.

2

Press 

User Management

The User Management screen opens listing the available user names and corresponding 
access levels and statuses.

3

Press 

Add

.

The Add New User screen opens.

4

In the User Name field, enter a user name for the new account. 
User names can contain a combination of letters and numbers, but numbers are only 
permitted at the end of the user name (e.g., abc123). Do not include special characters in the 
user name.

5

Next to 

Access Level

, select the access level (Standard or Advanced) for the new user. The 

default selection is 

Standard

.

See 

Table 4

 for a summary of the differences between the two access levels.

6

In the Password and Confirm Password fields, enter a password for the account.

7

In the Email address(es) field, enter one or more email addresses to associate with the user 
account.
Separate multiple email addresses with a space. You can enter up to five email addresses.
When the system needs to send an email alert to the user, it sends it to all email addresses 
listed in the Email address(es) field. 

8

Mark the check box labeled 

Email alert on run complete

 if you want the system to send an 

email alert to the user when it completes a protocol that was initiated by that user.

9

Mark the check box labeled 

Email alert on error occurs

 if you want the system to send an 

email alert to the user when it encounters an error during a protocol run that was initiated by 
that user.

10

Press 

OK

 to save the user account. 

The Add New User screen closes and you are returned to the User Management screen. The 
new user name appears in the list on the User Management screen.

Decontaminating the System

The Magnis NGS Prep System includes a UV tube that can be used to decontaminate the surfaces 
of the instrument deck. The 

quick cycle

 is a 30-minute decontamination. Agilent recommends 

running the quick cycle before each protocol run. The 2-hour 

extended cycle

 is recommended in 

the event of a spill or leak on the deck surfaces.

Instructions for running a quick cycle are provided below. For instructions on running an extended 
cycle, see the full-length 

Magnis NGS Prep System User Guide 

(publication K1007-90000).

WARNING

Do not look directly at the UV light while decontamination is in progress.

Содержание Magnis NGS Prep System

Страница 1: ...7 Log in to the system 7 Creating User Accounts 8 About user access levels 8 Add new user accounts 9 Decontaminating the System 9 Running Protocols 10 1 System prep 10 2 Reagent prep 11 3 Protocol se...

Страница 2: ...tem is designed for safe operation when it is used in the intended manner Using the system for purposes other than those intended may impair those safety protections Installation The system must be in...

Страница 3: ...not service the electrical components unless you are qualified to do so Fluids and Reagents Never process reagents not intended for use with the Magnis NGS Prep System Observe the relevant safety reg...

Страница 4: ...rument in high static environments ESD STM5 1 1998 Class 3B Environmental Requirements Operating temperature Keep the ambient temperature between 15 C and 25 C Operating humidity Keep the humidity lev...

Страница 5: ...eps Chiller module for reagent storage Tube holders for liquid reagents 6 total Platforms to hold tip boxes 4 Platform to hold Beads Buffers Plate Opening to tip waste bin LED indicator lights 2 Micro...

Страница 6: ...status of the instrument based on the color of the LED indicator lights that illuminate the entire plate filling area Micropipettor Barcode scanner UV tube Instrument deck LED indicator light LED indi...

Страница 7: ...the user name and password for your account into the provided fields The Agilent engineer or Agilent authorized service provider creates a user account during system installation that has advanced us...

Страница 8: ...of markers called teach points that are printed on the instrument deck Auto teaching ensures that during a run the micropipettor is accurately aligned with the tubes or wells at each deck position You...

Страница 9: ...s to send an email alert to the user it sends it to all email addresses listed in the Email address es field 8 Mark the check box labeled Email alert on run complete if you want the system to send an...

Страница 10: ...prep Prepare the Magnis NGS Prep System so that it is ready to run a protocol 1 Verify that the instrument deck is cleared of all labware from previous runs 2 Turn on the instrument and close the inst...

Страница 11: ...orms an IHC At completion of the IHC the Enter Run Info step opens 2 Complete the Run Setup steps and start the protocol run as instructed in the user manual for your Magnis Target Enrichment Kit The...

Страница 12: ...llect the final library samples i e the green library strip tube from the chiller module 3 Re seal the wells of the strip tube using a fresh foil seal strip then store at the temperature recommended i...

Страница 13: ...7 9770 option 3 4 4 Or send an email to techservices agilent com For Europe Middle East Africa and India Call 00800 345 600 toll free or 49 69 8679 7730 Or send an email to genomics_tech_europe agilen...

Отзывы: