![Agilent Technologies Magnis NGS Prep System Скачать руководство пользователя страница 2](http://html1.mh-extra.com/html/agilent-technologies/magnis-ngs-prep-system/magnis-ngs-prep-system_quick-reference-manual_2867774002.webp)
Before You Begin
2
Magnis NGS Prep System Quick Reference Guide
Before You Begin
Product
Description
The Magnis NGS Prep System is an automated liquid handling system for next generation
sequencing library preparation and/or target enrichment of human nucleic acid samples.
Principle of the
Procedure
The Agilent Magnis NGS Prep System is a liquid handler that offers start-to-finish automation of
library preparation and target enrichment protocols for next generation sequencing (NGS). The
starting material is fragmented genomic DNA (gDNA) purified from a cell or tissue sample, blood
sample, or formalin-fixed paraffin-embedded (FFPE) sample. The end result is a target-enriched
DNA library ready for sequencing.
The hardware components make up the instrument. See
a list of these components.
The instrument is controlled through the software component of the system, which is displayed
and operated through the LCD touchscreen.
Limitations of
Use
The Magnis NGS Prep System has been validated for use with Agilent Magnis NGS kits.
Materials
Provided
Safety Precautions
The Magnis NGS Prep System is designed for safe operation when it is used in the intended
manner. Using the system for purposes other than those intended may impair those safety
protections.
Installation
•
The system must be installed by an Agilent engineer or Agilent authorized service provider.
•
Do not attempt to manually lift the instrument. To move the instrument, use an automated
forklift or lift table that can lift at least 100 kg.
•
When adjusting the placement of the instrument on the laboratory bench, take care to avoid
pinching hazards on the instrument.
•
Retain the locking jigs that secure the gantry during shipment. Anytime the instrument is
moved, secure the gantry using the jigs.
Electrical
Standard electrical safety precautions should be applied, including the following:
•
Per North American and IEC requirements, install the instrument in a location where branch
circuit protection is available in the power network.
•
Install the instrument with the Agilent supplied power cord that is compatible with the electrical
outlets in your region. Do not substitute with a power cord from another source.
Table 1
Materials provided with the Magnis NGS Prep System
Materials
Magnis NGS Prep System instrument and preloaded touchscreen software
Power cord
Functional test certificate
USB disk with PDFs of the User Guide and Quick Reference Guide