Safety Precautions

Magnis NGS Prep System Quick Reference Guide

3

Install the instrument in a location where the power cord can be easily reached for quick 
disconnection of the power supply.

Install the instrument away from any flammable sources.

Make sure that the ventilation slots of the instrument are free to vent at all times. Maintain 

10 cm of cleared space on each side of the instrument and 18 cm of cleared space at the back 
of the instrument.

Connect the power cord to a wall output that provides 100–240 VAC, 1000 W.

Make sure that proper voltage is supplied before turning on the instrument for the first time.

Connect the instrument to a grounded socket. Do not operate the instrument from a power 
outlet that has no ground connection.

Do not connect the instrument to the same circuit as other devices with a high current draw 
(e.g., freezers, centrifuges). If possible, connect the instrument to an independent or dedicated 
AC circuit. 

Do not touch any switches or outlets with wet hands. 

Before disconnecting the power cord, turn off the instrument at both the power button on the 
front and the power switch on the back.

Unplug the instrument before you clean any major liquid spills and before servicing any of the 
electrical or internal components.

Do not operate the instrument in a hazardous or potentially explosive environment. 

Do not service the electrical components unless you are qualified to do so

.

Fluids and

Reagents

Never process reagents not intended for use with the Magnis NGS Prep System.

Observe the relevant safety regulations when handling pathogenic material, radioactive 
substances or other substances hazardous to health.

Do not submerge the instrument in any liquid.

Danger of

Ultraviolet (UV)

Light Exposure

Although exposure to UV light is minimal, the following precautions are still needed.

During decontamination of the instrument deck with UV light, do not look directly or indirectly 
at the UV light source. 

Always perform decontamination with the instrument door closed and locked. The instrument 
door is programmed to remain locked while the UV light is on.

Replacement UV tubes must be provided by Agilent and must be installed by an Agilent 
engineer or Agilent authorized service provider.

Danger of Burns

During protocol runs, the thermal block and other components of the thermal cycler module 
quickly attain temperatures of greater than 50°C. Keep the instrument door closed during runs. 
The instrument door is programmed to remain locked while protocol runs are in progress.

Use only Agilent materials (plates, adhesive seals, foils, mats) intended for use on the Magnis 
NGS Prep System. These materials are sufficiently temperature-stable (up to 120°C).

Equipment

Ratings

Pollution degree 2

Installation category II

Altitude 2000 m

Humidity 30 to 70%, non-condensing

Electrical supply 100–240 V, 50/60 Hz, 1000 W

Temperature 15°C to 25°C

For Indoor Use Only

Содержание Magnis NGS Prep System

Страница 1: ...7 Log in to the system 7 Creating User Accounts 8 About user access levels 8 Add new user accounts 9 Decontaminating the System 9 Running Protocols 10 1 System prep 10 2 Reagent prep 11 3 Protocol se...

Страница 2: ...tem is designed for safe operation when it is used in the intended manner Using the system for purposes other than those intended may impair those safety protections Installation The system must be in...

Страница 3: ...not service the electrical components unless you are qualified to do so Fluids and Reagents Never process reagents not intended for use with the Magnis NGS Prep System Observe the relevant safety reg...

Страница 4: ...rument in high static environments ESD STM5 1 1998 Class 3B Environmental Requirements Operating temperature Keep the ambient temperature between 15 C and 25 C Operating humidity Keep the humidity lev...

Страница 5: ...eps Chiller module for reagent storage Tube holders for liquid reagents 6 total Platforms to hold tip boxes 4 Platform to hold Beads Buffers Plate Opening to tip waste bin LED indicator lights 2 Micro...

Страница 6: ...status of the instrument based on the color of the LED indicator lights that illuminate the entire plate filling area Micropipettor Barcode scanner UV tube Instrument deck LED indicator light LED indi...

Страница 7: ...the user name and password for your account into the provided fields The Agilent engineer or Agilent authorized service provider creates a user account during system installation that has advanced us...

Страница 8: ...of markers called teach points that are printed on the instrument deck Auto teaching ensures that during a run the micropipettor is accurately aligned with the tubes or wells at each deck position You...

Страница 9: ...s to send an email alert to the user it sends it to all email addresses listed in the Email address es field 8 Mark the check box labeled Email alert on run complete if you want the system to send an...

Страница 10: ...prep Prepare the Magnis NGS Prep System so that it is ready to run a protocol 1 Verify that the instrument deck is cleared of all labware from previous runs 2 Turn on the instrument and close the inst...

Страница 11: ...orms an IHC At completion of the IHC the Enter Run Info step opens 2 Complete the Run Setup steps and start the protocol run as instructed in the user manual for your Magnis Target Enrichment Kit The...

Страница 12: ...llect the final library samples i e the green library strip tube from the chiller module 3 Re seal the wells of the strip tube using a fresh foil seal strip then store at the temperature recommended i...

Страница 13: ...7 9770 option 3 4 4 Or send an email to techservices agilent com For Europe Middle East Africa and India Call 00800 345 600 toll free or 49 69 8679 7730 Or send an email to genomics_tech_europe agilen...

Отзывы: