For the CT model only, check battery operation by disconnecting the AC power cord
from the AC power source.
___The leftmost LED on the power supply should go out after a few seconds. The other two
LEDs to the right of it should remain illuminated.
___The battery fuel gauge should display on the screen indicating the capacity of the
batteries. If the battery fuel gauge indicates that the batteries are discharged, reconnect the
AC power cord to recharge batteries or replace with fully charged batteries.
___If the Flat Panel display dies when AC power cord is removed, the batteries are either not
present or are discharged. Check to see if the batteries are present. If the batteries are not
present, install batteries. If batteries are present, reconnect the AC power cord to recharge
batteries or replace with fully charged batteries. see “Replacing Lead-Acid Batteries” on page
9-18 for information on installing batteries.
___If charged batteries are installed but the battery indicator does not appear on the display,
the mechanical contact that drives the indicator is faulty. See “Battery Problems” on page 14
or replace the power supply assembly to correct the fault. See “Removing the Power Supply
Assembly” on page -10 for information on replacing the power supply assembly.
0RQRFKURPH)ODW3DQHO
__If the
PRQRFKURPH
Flat Panel backlight fails to light, the source of the problem may be
the cable connecting the System Board to the Flat Panel Adapter Board, Flat Panel Adapter
Board, cable connecting Flat Panel Adapter Board to Flat Panel, or the backlight.
__If the monochrome screen is either all white or deep blue, and doesn’t change with
movement of the contrast control thumbwheel, the problem could be failure of the Flat Panel
contrast circuit on the Flat Panel Adapter Board, the contrast control potentiometer, or cable.
__If the monochrome screen can be adjusted in brightness by the contrast control, but there is
no display, then the System Board, Flat Panel assembly, or possibly cables could be
defective.
The backlight inverter circuitry is located on the monochrome Flat Panel Adapter Board.
127(
Содержание M1205A V24
Страница 1: ...3DUW 1R 0 3ULQWHG XQH 7KLUG GLWLRQ JLOHQW 0 9 DQG 9 6HUYLFH XLGH...
Страница 4: ......
Страница 12: ...12 Contents...
Страница 16: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 2YHUYLHZ...
Страница 40: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH QWURGXFLQJ WKH JLOHQW 9 9 6HULHV Introducing the Agilent V24 V26 Series...
Страница 214: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 5HSDLULQJ WKH JLOHQW 0RGHO 9 Repairing the Agilent Model V24...
Страница 251: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 5HSDLULQJ WKH JLOHQW 0RGHO 9 7 9 7 Repairing the Agilent Model V24CT V26CT...
Страница 252: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 5HSDLULQJ WKH JLOHQW 0RGHO 9 7 9 7 Repairing the Agilent Model V24CT V26CT...
Страница 301: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 5HSODFHPHQW 3DUWV Replacement Parts 17 4 3 7 12 5 6 26...
Страница 302: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 5HSODFHPHQW 3DUWV Replacement Parts 12 21 20 18 19 17 16 7 4 12 24...
Страница 308: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 5HSODFHPHQW 3DUWV Replacement Parts 9HUVLRQ 41 40 9 45 38 43 44 37 39 36 42 35...
Страница 315: ...JLOHQW 0 9 9 6HULHV 6HUYLFH XLGH 5HSODFHPHQW 3DUWV Replacement Parts 9HUVLRQ 41 40 9 45 38 43 44 37 39 36 42 35...
Страница 358: ...QGH...