Setting Up the Instrument
9
Synchronizing the Pattern
Now make sure that the error detector
automatically syncs to the data pattern:
With this setting, the error detector starts
synchronizing when the BER exceeds a
certain threshold value. You can see when
it has synchronized when the
indicator goes off.
Sync Loss
Setting the Sampling Point
Now set the optimum sampling point:
Open the
dialog box.
Pattern Sync
Setup
Make sure that
and
are selected.
Normal Sync Mode
Automatic Sync
1
2
Switch to the
Sampling Point Setup.
Press the
button
to measure the eye diagram
and to position the sampling
point at the center of the eye.
Auto Align
This usually takes
about 10 seconds.
The BER indicator in the top left
corner displays the current BER
at this sampling point which
should be 0 for this test.
1
2
If sync fails, the sampling point is
probably outside of the eye. You should proceed with
setting the sampling point anyway. As soon as the instrument
finds the eye, with the settings we just made, it will
automatically synchronize the patterns.