Preparing the Test

6

What Are We Testing?

The quickest way to familiarize yourself with the instrument is to set up a
short test. You can test the BER on a cable. All you need for this test is an
APC/RPC cable for the data port connection and an APC/RPC-SMA adapter
cable to connect the PG clock output to the ED clock input.

Setting up the Pattern Generator

First we need to set up the pattern generator.
To keep things simple, let's set up LVDS logic
levels for both the data and clock output.

Reset the instrument.

This sets all voltages
and parameters to
default values.

Now select the logic
family LVDS for the data
and clock port.

Check the values to
make sure you can’t
damage your device.

You can change the output
signal parameters according
to the needs of your device.

Let's increase the
signal's data offset
by 100 mV. Just tip
the offset field,

and turn the fine-tuning
knob until you have the
desired value.

1

The pattern generator's

output ports are always active.

Therefore, you must be sure that

your device can handle these

ranges before you connect it.

3

4

5

USB

Local

Preset

Switch to the
PG Setup.

2

Use the Up and Down buttons below

the knob to position the cursor.

Содержание J-BERT N4903A

Страница 1: ...mation on the J BERT check out the Online Help The Help also offers printable versions of the User Guide and the Programming Guide You only need a few minutes to get started with the J BERT This Getti...

Страница 2: ...specified in the operation instructions All Light Emitting Diodes LEDs used in this product are Class 1 LEDs as per IEC 60825 1 Environmental Conditions Verify that all safety precautions are taken Th...

Страница 3: ...m any single phase AC power source supplying 100 240 V in the frequency range from 47 63 Hz The maximum power consumption is 450 VA with all options installed The power supply automatically adapts to...

Страница 4: ...PS 2 Mouse green Serial Port 1 Parallel Port VGA Monitor Ethernet Port 1 2 Audio Line Output Jack Audio Microphone Input Jack Diagnostic LED Stack USB 2 0 Channel 0 3 The Agilent J BERT is available i...

Страница 5: ...values by hand or the large knob to adjust values You can also use the knobs at the bottom of the instrument to change certain frequently used values in run time Navigating the GUI The menu buttons al...

Страница 6: ...tput Reset the instrument This sets all voltages and parameters to default values Now select the logic family LVDS for the data and clock port Check the values to make sure you can t damage your devic...

Страница 7: ...e used for the test Select the Pattern Press the button to switch to the pattern editor Pattern Setup Open the dialog box Pattern Select Select the pattern type 2 n 1 PRBS Select the pattern size 2 31...

Страница 8: ...etector uses the clock from the pattern generator 2 PG Clock Out PG Data Out ED Data In ED Clock In Press the button twice to switch to the error detector setup PG ED Setup Click the button Edit Set t...

Страница 9: ...e sure that and are selected Normal Sync Mode Automatic Sync 1 2 Switch to the Sampling Point Setup Press the button to measure the eye diagram and to position the sampling point at the center of the...

Страница 10: ...to move the sampling point horizontally Data In Delay You can see how the BER increases when measuring at the edge of the eye 1 2 Viewing Eye Results Now let s check the resulting eye Press the butto...

Страница 11: ...le be used for testing error correction algorithms Select because we do not need a measurement log file now No Logging Select a of 10 seconds as the measurement period Time 3 4 5 Press the button to s...

Страница 12: ...ument The software will automatically recognize the new component and provide the Interface Channel Icon and according setup functions in the window Jitter Perform the regular shutdown procedure provi...

Отзывы: