156
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
5
Common (*) Commands
selected and a common command is received by the instrument, the
instrument remains in the selected subsystem. For example, if the program
message ":ACQuire:TYPE AVERage; *CLS; COUNt 256" is received by the
instrument, the instrument sets the acquire type, then clears the status
information and sets the average count.
In contrast, if a root level command or some other subsystem command is
within the program message, you must re- enter the original subsystem
after the command. For example, the program message ":ACQuire:TYPE
AVERage; :AUToscale; :ACQuire:COUNt 256" sets the acquire type,
completes the autoscale, then sets the acquire count. In this example,
:ACQuire must be sent again after the :AUToscale command in order to
re- enter the ACQuire subsystem and set the count.
N O T E
Each of the status registers has an enable (mask) register. By setting the bits in the enable
register, you can select the status information you want to use.
Содержание InfiniiVision 3000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Страница 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Страница 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Страница 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Страница 152: ...152 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 4 Commands Quick Reference...
Страница 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Страница 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Страница 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Страница 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Страница 306: ...306 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 14 EXTernal Trigger Commands...
Страница 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Страница 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Страница 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Страница 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Страница 466: ...466 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 20 MEASure Power Commands...
Страница 502: ...502 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 22 POD Commands POD1 DISP 0 THR 1 40E 00...
Страница 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Страница 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Страница 1046: ...1046 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 34 Obsolete and Discontinued Commands...
Страница 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Страница 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Страница 1086: ...1086 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 37 Synchronizing Acquisitions...
Страница 1216: ...1216 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 39 Programming Examples...
Страница 1231: ...Index Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 1231 zoomed window horizontal scale 841...
Страница 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...