Obsolete and Discontinued Commands
34
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
1025
:MEASure:TVOLt
Query Syntax
:MEASure:TVOLt? <value>, [<slope>]<occurrence>[,<source>]
<value> ::= the voltage level that the waveform must cross.
<slope> ::= direction of the waveform.
A rising slope is indicated by
a plus sign (+).
A falling edge is indicated by a minus
sign (-).
<occurrence> ::= the transition to be reported.
If the occurrence
number is one, the first crossing is reported.
If
the number is two, the second crossing is reported,
etc.
<source> ::= {<digital channels> | CHANnel<n> | FUNCtion | MATH}
<digital channels> ::= {DIGital<d>} for the MSO models
<n> ::= 1 to (# analog channels) in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
When the :MEASure:TVOLt? query is sent, the displayed signal is searched
for the specified voltage level and transition. The time interval between the
trigger event and this defined occurrence is returned as the response to
the query.
The specified voltage can be negative or positive. To specify a negative
voltage, use a minus sign (- ). The sign of the slope selects a rising (+) or
falling (- ) edge. If no sign is specified for the slope, it is assumed to be
the rising edge.
The magnitude of the occurrence defines the occurrence to be reported.
For example, +3 returns the time for the third time the waveform crosses
the specified voltage level in the positive direction. Once this voltage
crossing is found, the oscilloscope reports the time at that crossing in
seconds, with the trigger point (time zero) as the reference.
If the specified crossing cannot be found, the oscilloscope reports
+9.9E+37. This value is returned if the waveform does not cross the
specified voltage, or if the waveform does not cross the specified voltage
for the specified number of times in the direction specified.
If the optional source parameter is specified, the current source is
modified.
Return Format
<value><NL>
N O T E
The :MEASure:TVOLt command is obsolete and is provided for backward compatibility to
previous oscilloscopes. Use the :MEASure:TVALue command (see
).
Содержание InfiniiVision 3000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Страница 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Страница 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Страница 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Страница 152: ...152 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 4 Commands Quick Reference...
Страница 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Страница 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Страница 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Страница 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Страница 306: ...306 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 14 EXTernal Trigger Commands...
Страница 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Страница 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Страница 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Страница 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Страница 466: ...466 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 20 MEASure Power Commands...
Страница 502: ...502 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 22 POD Commands POD1 DISP 0 THR 1 40E 00...
Страница 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Страница 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Страница 1046: ...1046 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 34 Obsolete and Discontinued Commands...
Страница 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Страница 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Страница 1086: ...1086 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 37 Synchronizing Acquisitions...
Страница 1216: ...1216 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 39 Programming Examples...
Страница 1231: ...Index Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 1231 zoomed window horizontal scale 841...
Страница 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...