938
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
31
:WAVeform Commands
:WAVeform:POINts:MODE
Command Syntax
:WAVeform:POINts:MODE <points_mode>
<points_mode> ::= {NORMal | MAXimum | RAW}
The :WAVeform:POINts:MODE command sets the data record to be
transferred with the :WAVeform:DATA? query.
For the analog or digital sources, there are two different records that can
be transferred:
•
The first is the raw acquisition record. The maximum number of points
available in this record is returned by the :ACQuire:POINts? query. The
raw acquisition record can only be transferred when the oscilloscope is
not running and can only be retrieved from the analog or digital
sources.
•
The second is referred to as the
measurement record
and is a
62,500- point (maximum) representation of the raw acquisition record.
The measurement record can be retrieved from any source.
If the <points_mode> is NORMal the measurement record is retrieved.
If the <points_mode> is RAW, the raw acquisition record is used. Under
some conditions, such as when the oscilloscope is running, this data
record is unavailable.
If the <points_mode> is MAXimum, whichever record contains the
maximum amount of points is used. Usually, this is the raw acquisition
record. But, if the raw acquisition record is unavailable (for example,
when the oscilloscope is running), the measurement record may have more
data. If data is being retrieved as the oscilloscope is stopped and as the
data displayed is changing, the data being retrieved can switch between
the measurement and raw acquisition records.
Considerations
for MAXimum or
RAW data
retrieval
•
The instrument must be stopped (see the :STOP command (see
) or the :DIGitize command (see
) in the root
subsystem) in order to return more than the
measurement record
.
•
:TIMebase:MODE must be set to MAIN.
•
:ACQuire:TYPE must be set to NORMal, AVERage, or HRESolution. If
AVERage, :ACQuire:COUNt must be set to 1 in order to return more
than the
measurement record
.
•
MAXimum or RAW will allow up to 4,000,000 points to be returned. The
number of points returned will vary as the instrument's configuration is
changed. Use the :WAVeform:POINts? MAXimum query to determine the
maximum number of points that can be retrieved at the current
settings.
Query Syntax
:WAVeform:POINts:MODE?
Содержание InfiniiVision 3000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Страница 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Страница 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Страница 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Страница 152: ...152 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 4 Commands Quick Reference...
Страница 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Страница 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Страница 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Страница 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Страница 306: ...306 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 14 EXTernal Trigger Commands...
Страница 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Страница 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Страница 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Страница 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Страница 466: ...466 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 20 MEASure Power Commands...
Страница 502: ...502 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 22 POD Commands POD1 DISP 0 THR 1 40E 00...
Страница 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Страница 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Страница 1046: ...1046 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 34 Obsolete and Discontinued Commands...
Страница 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Страница 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Страница 1086: ...1086 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 37 Synchronizing Acquisitions...
Страница 1216: ...1216 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 39 Programming Examples...
Страница 1231: ...Index Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 1231 zoomed window horizontal scale 841...
Страница 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...