![Agilent Technologies HP E5574A Скачать руководство пользователя страница 170](http://html1.mh-extra.com/html/agilent-technologies/hp-e5574a/hp-e5574a_user-manual_2868134170.webp)
168
Appendix A. Installation and Maintenance
AC Line Power Supply Requirements
The AC power requirements are summarized on the rear panel of
the instrument.
Replacing the Fuse
There is one fuse in this instrument. This is a T1.5A/250V (time-
lag) (HP Part No. 2110-0304). The fuse holder is at the rear of the
instrument, beside the line power connector.
How to Replace the Fuse
1
Release the fuse holder: use the blade of a flat-headed
screwdriver to depress the catch at the side of the holder and then
pull the holder out a little.
Figure A-2
Rear Panel Markings
ISM GROUP 1 CLASS B
Figure A-3
Releasing the Fuse Holder
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание HP E5574A
Страница 11: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 75: ...5 5 Measuring Power Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 177: ...B B Accessories Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 185: ...C C Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 186: ...184 Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 195: ...D D Performance Test Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 223: ...E E Cleaning Procedure Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 243: ...F F Error Messages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...