![Agilent Technologies HP 5890 II Plus Series Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/agilent-technologies/hp-5890-ii-plus-series/hp-5890-ii-plus-series_manual_2868131009.webp)
Site Preparation
Power considerations
8
Power considerations
Line voltage
Your gas chromatograph operates from one of these voltage supplies,
depending on the power option ordered. These options and the power
cords furnished are described on the next page.
C
120 VAC single phase (+5, -10%), 48 to 66 Hz, 2200 VA max.
(requires a 20 amp dedicated line)
C
200 VAC split phase (+5, -10%), 48 to 66 Hz, 2200 VA max.
(requires a 15 amp dedicated line)
C
220/240 VAC single phase (+5, -10%), 48 to 66 Hz, 2200 VA max.
(requires a 15 amp dedicated line)
C
220/240 VAC split phase (+5, -10%), 48 to 66 Hz, 2200 VA max.
(requires a 15 amp dedicated line)
Power line cord
To protect personnel, the instrument panels and cabinet are grounded
through the three•conductorpower line cord in accordance with
International Electrotechnical Commission (IEC) requirements.
The three•conductorpower line cord, when plugged into a properly
grounded receptacle, provides power and grounds the instrument. A
properly grounded receptacle has its ground contact connected to a
suitable earth ground. Proper receptacle grounding should be verified.
The power options for the HP 5890, and the electrical power and the line
cord termination for each, are shown here.
Содержание HP 5890 II Plus Series
Страница 1: ...Chemical Analysis Group 5890 Gas Chromatograph 5890 Series II Site Prep Info Document A15282 ...
Страница 2: ...Site Prep Installation Manual HP 5890 Series II and HP 5890 Series II Plus ...
Страница 6: ...1 Site Preparation ...
Страница 19: ...This page intentionally left blank ...
Страница 20: ...2 Installation ...