Multimode Source for 6410 Triple Quad LC/MS User Guide
39
Reference
3
Benefits of a multimode source
•
Permits analysis of fast-eluting peaks
•
Operation at flow rates used for high-throughput analysis without splitting
•
Operation at lower flow rates for higher-sensitivity analysis
The design uses a single nebulizer through which the LC effluent flows. You do
not need to switch back and forth between APCI and ESI modes. While this
option will be available, the great value of the source is its ability to produce
APCI and ESI generated ions simultaneously. This ability increases the
effective duty cycle of the mass spectrometer. Another key feature is the much
greater operational flow rate that is available in this design.
Mixed mode operation is generally a balance between optimal ESI and APCI
conditions. Changing the vaporizer temperature, the nebulizer pressure, and
the corona current alters the balance between the ionization modes.
Ionization Mode
Polarity
APCI
positive
APCI
negative
ESI
positive
ESI
negative
Multimode
positive
Multimode
negative
Parameter adjustment
APCI response in mixed mode
generally
ESI response in mixed mode
generally
Higher vaporizer temperature
Increases
Stays the same
Higher nebulizer pressure
Decreases
Increases
Higher corona current
Increases
Decreases