Site Requirements and Specifications
Performance Specifications
Performance Specifications
Table 5
Performance Specifications Agilent 1200 Infinity Autosampler (G1329A). Valid
when standard 100 µl metering head installed.
Type
Specification
Pressure
Operating range 0 – 40 MPa (0 – 400 bar, 0 – 5900 psi)
GLP features
Early maintenance feedback (EMF), electronic records of
maintenance and errors
Communications
Controller-area network (CAN). GPIB (IEEE-448), RS232C,
APG-remote standard, optional four external contact closures and
BCD vial number output
Safety features
Leak detection and safe leak handling, low voltages in maintenance
areas, error detection and display
Injection range
0.1 – 100 µl in 0.1 µl increments Up to 1500 µl with multiple draw
(hardware modification required)
Replicate injections
1 – 99 from one vial
Precision
< 0.25% RSD from 5 – 100 µl, < 1% RSD 1 – 5 µl
variable volume
Minimum sample volume
1 µl from 5 µl sample in 100 µl microvial, or 1 µl from 10 µl sample in
300 µl microvial
Carryover
Typically < 0.1%, < 0.05% with external needle cleaning
Sample viscosity range
0.2 – 50 cp
Replicate injections per vial
1 – 99
Sample capacity
100 × 2-ml vials in 1 tray
40 × 2-ml vials in ½ tray
15 × 6-ml vials in ½ tray (Agilent vials only)
Injection cycle time
Typically 50 s depending on draw speed and injection volume
Содержание G1329A
Страница 1: ...Agilent 1260 Infinity Preparative Autosampler User Manual...
Страница 6: ...6 Contents...
Страница 22: ...22 1 Introduction to the Autosampler Agilent 1200 Infinity Interfaces...
Страница 32: ...32 2 Site Requirements and Specifications Performance Specifications...
Страница 52: ...52 3 Installing the Autosampler Transporting the Autosampler...
Страница 74: ...74 6 Troubleshooting and Diagnostics Overview of the Sampler s Indicators and Test Functions...
Страница 106: ...106 8 Parts and Materials for Maintenance Main Assemblies Main Assemblies Figure 23 Autosampler Main Assemblies...
Страница 109: ...109 Parts and Materials for Maintenance 8 Analytical Head Assembly Figure 24 Analytical Head Assembly cdi h dlc...
Страница 136: ...136 9 Identifying Cables RS 232 Cables...
Страница 149: ...149 Index...