![Agilent Technologies G1329A Скачать руководство пользователя страница 130](http://html1.mh-extra.com/html/agilent-technologies/g1329a/g1329a_user-manual_2867725130.webp)
130
9
Identifying Cables
BCD Cables
BCD Cables
One end of these cables provides a 15-pin BCD connector to be connected to
the Agilent 1200 Infinity modules. The other end depends on the instrument
to be connected to
Agilent Modules to General Purpose
Connector G1351-81600
Wire Color
Pin Agilent
module
Signal Name
BCD Digit
Green
1
BCD 5
20
Violet
2
BCD 7
80
Blue
3
BCD 6
40
Yellow
4
BCD 4
10
Black
5
BCD 0
1
Orange
6
BCD 3
8
Red
7 BCD
2
4
Brown
8
BCD 1
2
Gray
9
Digital ground
Gray
Gray/pink
10
BCD 11
800
Red/blue
11
BCD 10
400
White/green
12
BCD 9
200
Brown/green
13
BCD 8
100
not connected
14
not connected
15
+ 5 V
Low
Содержание G1329A
Страница 1: ...Agilent 1260 Infinity Preparative Autosampler User Manual...
Страница 6: ...6 Contents...
Страница 22: ...22 1 Introduction to the Autosampler Agilent 1200 Infinity Interfaces...
Страница 32: ...32 2 Site Requirements and Specifications Performance Specifications...
Страница 52: ...52 3 Installing the Autosampler Transporting the Autosampler...
Страница 74: ...74 6 Troubleshooting and Diagnostics Overview of the Sampler s Indicators and Test Functions...
Страница 106: ...106 8 Parts and Materials for Maintenance Main Assemblies Main Assemblies Figure 23 Autosampler Main Assemblies...
Страница 109: ...109 Parts and Materials for Maintenance 8 Analytical Head Assembly Figure 24 Analytical Head Assembly cdi h dlc...
Страница 136: ...136 9 Identifying Cables RS 232 Cables...
Страница 149: ...149 Index...