Installation Note E8364-90023
35
7. Connect the ribbon cable (item
④
) and place the mixer bracket, with all four mixers
attached, into the analyzer. Observe the
CAUTION
below.
CAUTION
Be careful not to damage the center pins of the semirigid cables. Some flexing of
the cables is necessary when installing the assembly. Do not over-bend them.
8. Using a T-10 TORX driver, install the four mounting screws (item
③
).
9. Using a 5/16-inch torque wrench (set to 10-inch lbs), reconnect the four semirigid cables
(item
①
) and the four flexible cables (item
②
) shown. Observe the
CAUTION
above. The
remaining cables will be installed later.
Figure 23
A27, A28, A29, and A30 Mixers Installation