Installation Note E8364-90023
11
Test Equipment Required for Post-Upgrade Adjustments
Step 1. Obtain a Keyword and Verify the Information
Follow the instructions on the Option Entitlement Certificate supplied to obtain a license key
for installation of this upgrade. Refer to
“License Key Redemption” on page 4
.
Verify that the model number, serial number, and option number information on the license
key match those of the instrument on which this upgrade will be installed.
If the model number, serial number, or option number do not match those on your license key,
you will not be able to install the option. If this is the case, contact Agilent for assistance
before beginning the installation of this upgrade. Refer to
“Contacting Agilent” on page 3
.
Once the license key has been received and the information verified, you can proceed with the
installation at step 2.
Description
Agilent Part Number
Power meter
E4418B/E4419B
Power sensor, Type-N (100 kHz–4.2 GHz)
8482A
Power sensor, 2.4 mm (4.2 GHz–50 GHz)
8487A
Power sensor, V-Band (50 GHz–67 GHz)
a
a. This sensor and adapter must be calibrated together by the Agilent factory. The data supplied
are only valid as long as the sensor and adapter remain connected.
V8486A
Adapter, WR-15 to 1.85 mm (m)
V281B
Adapter, 1.85 mm (f) to 1.85 mm (f)
b
b. The 1.85 mm connector has the same ruggedness and is compatible with the 2.4 mm connec-
tor. 1.85 mm and 2.4 mm connectors and adapters may be used interchangeably.
85058-60114
Adapter, 3.5 mm (m) to 2.4 mm (f)
11901D
Adapter, 2.4 mm (f) to 2.4 mm (f)
11900B
Adapter, Type-N (f) to 2.4 mm (f)
11903B
Adapter, SMA (m) to SMA (f), right angle
1250-1249
Test cable, 2.4 mm (f) to 2.4 mm (f)
Test cable, 1.85 mm (f) to 1.85 mm (f)
85133C
N4697E
5/16-inch, open-end torque wrench (set to 10 in-lbs)
8710-1765
1/4-inch, open-end wrench, thin profile
8710-0510