Installation Note E8364-90019
Description of Option 014
An Option 014 analyzer can be configured to measure high-power devices and devices that
require high dynamic range.
For a high-power measurement, external amplifiers and high power attenuators or isolators
can be added to complete the test setup. In this configuration, test-port output-power up to
1 Watt (+30 dBm) can be applied to the device under test (DUT). Additionally, there is an
external reference input that allows the external amplifier’s frequency response and drift to
be ratioed out.
For high dynamic range measurements, front panel jumpers are configured to reverse the
signal path through one of the couplers. This allows for a 15 dB improvement in transmitted
signal sensitivity in one direction only. These jumpers can also be configured on both ports,
allowing the user to perform high dynamic range measurements in both the forward and
reverse directions.