Installation Note E8364-90019
17
Analyzers WITH Options UNL AND 016
for the following procedure. The new parts referenced in this procedure are
listed in
Install these cables
in the order listed
. Use a 5/16-inch torque wrench set to 10 in-lbs.
• W71 E8364-20126 REFERENCE 2 RCVR R2 IN to A29 channel R2 mixer
• W70 E8364-20151 REFERENCE 1 RCVR R1 IN to A28 channel R1 mixer
• W68 E8364-20044 A26 test port 1 coupler to PORT 2 CPLR ARM
• W50 E8364-20130 PORT 2 RCVR B IN to A44 step attenuator
• W82 E8364-20054 A37 step attenuator to PORT 2 SOURCE OUT
• W84 E8364-20040 PORT 2 CPLR THRU to A39 bias tee
• W66 E8364-20048 Channel R2 detector to REFERENCE 2 SOURCE OUT
• W67 E8364-20043 A25 test port 1 coupler to PORT 1 CPLR ARM
• W49 E8364-20129 PORT 1 RCVR A IN to A43 step attenuator
• W81 E8364-20053 A36 step attenuator to PORT 1 SOURCE OUT
• W83 E8364-20039 PORT 1 CPLR THRU to A38 bias tee
• W65 E8364-20047 Channel R1 detector to REFERENCE 1 SOURCE OUT
Figure 10 Cable Installation, Analyzers with Options UNL and 016